Zoll X Series Defibrillator

  • Zoll X Series Defibrillator

Weighing under 12 pounds, the Zoll X Series Defibrillator is full-featured and designed to be effective for all patients, including neonates. Transporting a critically ill patient through the hospital requires a lightweight device with extensive monitoring capabilities. And that’s exactly what you get with the Zoll Defibrillator X Series, plus defibrillation and real-time CPR feedback in one compact device.

Description

Zoll X Series Defibrillator Features

Weighing under 12 pounds, the Zoll X Series Defibrillator is full-featured and designed to be effective for all patients, including neonates.

  • Capnography
  • Pulse oximetry
  • Three invasive pressures
  • Two temperature channels
  • Non-invasive blood pressure

Zoll X Series Defibrillator Specifications

Dimensions and Weight

  • Height with Handle: 10.4 in (26.4 cm)
  • Height without the Handle: 8.75 in (22.2 cm)
  • Width: 8.9 in (22.6 cm)
  • Depth: 7.9 in (20.1 cm)
  • Weight without the battery and Paper: 10.6 lbs
  • Weight with the battery and paper: 11.7 lbs

Operating

  • Temperature: 0to 50°CM
  • Humidity: 15 to 95% RH (non-condensing)
  • Vibration
    • EN ISO 9919 (per IEC 60068-2-64)
    • RTCA/DO-160G (multiple helicopter frequencies)
    • EN 1789 for ambulance
  • Shock: IEC 60068-2-27, 100g, 6ms half sine
  • Drop:
    • EN 1789, 30” functional drop
    • IEC 60601-1, Tested at 2 meters
  • Altitude: -170 M to 4572 M (-557 feet to 15,000 feet)

Temperature and Storage

  • Temperature: -30 to 70°C
  • Humidity: 15 to 95% RH (non-condensing)
  • Atmospheric pressure: 572 mbar to 1034 mbar
  • Shock/vibration: MIL-STD 810G, Method 514.6, 4.4.2, Procedure II
  • Safety Classification: Class 1 and internal power per IEC/EN 60601-1

AC Auxiliary Operating Power

  • Input
    • 100-240V 50-60 Hz, 2A
    • 100-115V 400 Hz, 2A
  • Output: 14.5V 4.15A 80W (peak)

Battery

  • Type: Rechargeable Lithium-Ion, 11.1Vdc, 6.6 Ah, 73Wh
  • Capacity:
    • At least 6 hours of continuous monitoring of ECG, SpO2, CO2, three Invasive Pressure channels, and 2 channels of Temperature, with NIBP measurements every 15 minutes and 10 200 J shocks (display set to 30%).
    • At least 100 discharges at maximum shock energy (200 joules).
    • At least 3.5 hours pacing, with ECG, SPO2, CO2, three Invasive Pressures, temperature, NIBP every 15 minutes, and pacing at 180 ppm, and 140 mA.
    • At least 300 discharges at maximum shock energy (200 joules) with no parameters and 70% brightness.
    • At least 10 discharges at maximum shock setting (200 joules) after a Low Battery indication.
  • Battery Indicators: 5 Battery capacity LED indicators, Fault indicator, Recalibration indicator
  • Recharge Rate: 100% in 4 hours, when initiated at Low Battery indication.

Monitor / Display

  • Size: 6.5 in (16.56 cm) diagonal
  • Input: 3-lead, 5-lead, or 12-lead patient cable, paddles, or multifunction electrodes
  • Type: Color LCD, 640 x 480 pixels, 800 MCD
  • Sweep Speed (user selectable): 25 mm/sec or 50 mm/se
  • Lead Selections: Paddles (Pads), I, II, III, AVR, AVL, AVF, V1-6
  • Frequency Response (user selectable)
    • 0.67 – 20 Hz Limited mode
    • 0.67 – 40 Hz Monitor mode
    • 0.25 – 40 Hz Filtered diagnostic mode
    • 0.05 – 150 Hz Diagnostic mode

Defibrillator

  • Waveform: Rectilinear Biphasic
  • Energy Selections: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 70, 85, 100, 120, 150, 200 joules
  • Energy Display: Display indicates selected and delivered energy
  • Charge Time: Less than 7 seconds with a new, fully charged battery
  • Charge Controls: Front panel and on apex paddle
  • Pads/Paddles: Hands-free pads and external adult and pediatric paddles. Adult paddles slide off to expose pediatric paddles.
  • Synchronized Mode: Synchronizes defibrillator pulse to patient’s R wave. “SYNC” message displayed on monitor and markers shown on both monitor and recorded ECG. Meets Clause 104 of AAMI DF80.

Pacer

  • Type: External transcutaneous pacing
  • Pacer Rate: 30 to 180 BPM ± 1.5%
  • Output Current: 0 to 140 mA ± 5% or 5 mA (whichever is greater)
  • Modes: Demand and Fixed