Zoll M Series CCT Defibrillator
The Zoll M Series CCT Defibrillator is designed with several user-selectable features that allow operators to configure the device according to their protocols and local requirements. With its straightforward controls and ZOLL Uniform Operating System, the M Series CCT reduces staff training and minimizes operator confusion. Thanks to the consistency built into each and every defibrillator unit, you train once and use it everywhere. Standardization has never been so easy.
Description
Zoll M Series CCT Defibrillator Features
The Zoll M Series CCT Defibrillator is designed with several user-selectable features that allow operators to configure the device according to their protocols and local requirements. With its straightforward controls and ZOLL Uniform Operating System, the M Series CCT reduces staff training and minimizes operator confusion. Thanks to the consistency built into each and every defibrillator unit, you train once and use it everywhere. Standardization has never been so easy.
- Superior defibrillation
- Superior pacing
- Advisory capabilities
- Transport-ready design
- Easy-to-read display
- Full range of parameters for full range of patients
- Flexible color display
- Quick, easy connections
Zoll M Series CCT Defibrillator Specifications
Dimensions and Weight
- With NIBP, IBP, Temp:
- Height: 10.2 in (25.9 cm)
- Width: 10.3 in (26.2 cm)
- Depth: 8.7 in (22.1 cm)
- Without NIBP, IBP, Temp:
- Height: 8.6 in (21.8 cm)
- Width: 10.3 in (26.2 cm)
- Depth: 8.7in (22.1 cm)
- Weight:
- 17.2 lbs ( 7.8 kg) with multi-function cable and battery
- 19.2 lbs (8.71 kg) with above and Paddles
AC Power
- 100-120 V, 50/60 Hz; 220-240 V, 50 Hz; 220 VA
Display Specifications
- Screen type: Active Matrix Color LCD
- Screen Size: 6.5 in (16.51 cm) Diagonal
- Number of Pixels: 640 x 480
- Sweep Speed: 5 mm/s
- Video Output: Industry standard VGA, 640 x 480, 60 Hz (OPTIONAL)
Battery Specifications (XL Battery Pack)
- Type: Rechargable, sealed lead acid
- Weight: 3.7 lbs (1.68 kg)
- Voltage: 2 volts/cell; 5 cells wired in series
- Recharge Time: 7.2 hours or less with integral charger
- Operating Time;For a new, fully charged XL battery pack at 20°C: 60 defibrillator discharges at maximum energy (200 J), or 2.5 hours minimum of continuous ECG and SpO2 monitoring, or 1.5 hours of continuous ECG, SpO2 , EtCO2 , IBP, and Temperature monitoring/pacing at 60 mA, 70 beats per minute.
Environmental Specifications
- Operating Temperature: 32° to 122° F (0° to 50° C)
- Storage and Shipping Temperature: -4° to 140° F (-20° to 60° C)
- Humidity: 5 to 95% relative humidity, noncondensing
- Vibration: Mil Std 810 E, Minimum Integrity Test
- Shock: IEC 68-2-27, 50g 6mS half sine
- Operating Pressure: 594 to 1060 millibars
Defibrillator
Waveform
- ZOLL Rectilinear Biphasic.
- Energy Selection: Selectable at 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 75, 100, 120, 150, 200 joules. (Delivered into a 50-ohm load.)
- Charge Time: Less than 6 seconds with a new fully charged battery (first 15 charges to 200J.)
- Depleted batteries will result in a longer defibrillator charge time.
- Energy Display: Monitor display indicates both selected and delivered energy.
Synchronized Mode
- Synchronizes defibrillator pulse to the patient’s R-wave.
- “SYNC” message displayed on the monitor.
- Marker on display and recorder paper identifies R-wave discharge point.
- Advisory Function: Single analysis or programmable auto re-analyze x3 with programmable auto energy level selection, screen prompts, and voice prompts.
- Charge Controls: Control on apex paddle and on device front panel.
- Paddles: External anterior/anterior adult and pediatric.
- Adult paddles slide off to expose pediatric paddles.
- Multi-Function Electrode (MFE) Pads: Specifically designed adult anterior/posterior pre-gelled ZOLL MFE Pads, and Multi-Function stat•padz™ packaged in pairs.
- Built-In Defibrillator Tester: Tests defibrillator energy output and continuity of universal cable and paddles; documented on PCMCIA card and strip chart.
Defibrillation Advisory
- Evaluates electrode connection and patient ECG to determine if defibrillation is required.
- Shockable Rhythms: Ventricular fibrillation with amplitude > 100 μV and wide complex ventricular tachycardia with rates greater than 150 bpm.
- Multi-Function Electrode Impedance Measurement.
Range
- 0-250 ohms
Display
Screen Type
- Color LCD.
- Screen Size: 6.5 inches (16.5 cm) diagonally.
- Sweep Speed: 25 mm/sec.
Viewing Time
- 4 seconds
- Traces: 3
- Information:
- Heart Rate
- Lead/Pads
- Alarm On/Off
- Advisory Functions and Prompts
- Defibrillator Test Function
- Error Corrections and Faults
- Code Markers
- Alarm Selection and Limits
- Delivered Energy
- SpO2
- Pacer Functions
- EtCO2
- NIBP
- Invasive Pressures (2)
- Temperature (2)
ECG Monitoring
Patient Connection
- 3-lead ECG cable, 5-lead ECG cable, 12 lead cable, paddles or MFE Pads.
- Input selection on the front panel.
- Input Protection: Fully defibrillator protected.
- Special circuit prevents distortion of ECG by pacer pulse. (Pacer version only.)
- Implanted Pacemaker Spike Display: Dedicated circuitry detects most implanted pacemaker spikes and provides a standard display marker of the spike on ECG trace.
- Bandwidth: 0.5-40 Hz (-3 dB) standard/0.05-150 Hz diagnostic.
- Lead Selection: Displayed on the monitor.
- ECG Size: 0.5, 1, 1.5, 2, 3 cm/mV – display on monitor.
- Heart Rate: Digital display 0-300 bpm ±5%.
- Heart Rate Alarm: On/Off displayed on the monitor.
- User-selectable: Tachycardia 60- 280 bpm, Bradycardia 20-100 bpm
Additional information
Brand | Zoll |
---|