Welch Allyn CP 150 EKG Machines
The Welch Allyn CP 150 EKG Machines with optional spirometry combines the best traditional box ECG features like full-size printouts and clinical decisions with a simple touchscreen interface, fast one-button operation, and flexible connectivity options to help you keep your staff productive and your tests accurate. And with the optional spirometry module, the CP 150 even lets you perform two reimbursable tests with just one device. The end result is a device that will help your facility save time, improve workflow, and efficiently manage patient information.
Description
Welch Allyn CP 150 EKG Machines Features
The Welch Allyn CP 150 EKG Machines with optional spirometry combines the best traditional box ECG features like full-size printouts and clinical decisions with a simple touchscreen interface
- 7” color touchscreen
- Battery
- Optional spirometry
- 100 test storage
Welch Allyn CP 150 EKG Machines Specifications
Dimensions and Weight
- Length: 380.9 mm (15 in)
- Height: 358.1 mm (14.1 in)
- Width: 136.2 mm (5.4in)
- Weight: 5.3 kg (11.7 kg) with the battery
Display
- Type: TFT, 18 cm (7 in) color touchscreen
- Resolution: WVGA, 800 x 480
Printing
- Thermal Paper: Z-fold 21 cm (8.25 in) x 28 cm (11 in) x 200 sheets
- Thermal Paper (internal): Computer-controlled dot array, 8 dots/mm
- Thermal chart paper speed: 10, 25 or 50 mm
- GainSettings:
- Auto ECGs: 2.5, 5, 10 or 20 mm/mV, AUTO
- Rhythm ECGs: 10, 25 or 50 mm/mV
Lead Configurations
- Standard, Cabera
Interpretation Algorithm
- MEANS (Modular ECG Analysis System)
Environmental Operating Conditions
- Temperature 10° – 40° C (50° – 104° F)
- Relative humidity 15 – 95% noncondensing (30 – 70% for printing)
- Atmospheric air-pressure limits 700 – 1060 hPa (525 – 795 mmHg)
Environmental Storage Conditions
- Temperature -20° – 50° C (-4° – 122° F)
- Relative humidity 15 – 95% noncondensing
- Atmospheric air-pressure limits 700 – 1060 hPa
Other Specifications
- Input impedance >2.5 MΩ
- Patient leakage <10 μA (normal condition), <50 μA (single fault condition)
- Dynamic range AC differential +/-5 mV, DC offset +/-300 mV
- Resolution 2.5 μV
- A/D conversion 16 bits
- Heart rate range 30 to 250 bpm
- ECG storage (in test directory) At least 100 ECG tests
- Frequency range 0.3 to 150 Hz
- Digital sampling rate 8000 samples/second/channel for data acquisition 500/1000 samples/second/channel for adult/pediatric ECG data analysis and storage
- Pacemaker detection ANSI/AAMI EC11
- Lead detection Lead off and noise
- Power requirement Universal AC power supply ~110–240 V, ~50/60 Hz, 1.5 A maximum
- AC fuses Time-lag type, 2.0-amp 250-V rating, Little fuse 0218002P or equivalent