ValleyLab Force FX and FXc Electrosurgical Units
The ValleyLab Force FX and FXc Electrosurgical Units are Digital Electrosurgical Units. The Force FXc is a High Power Digital Electrosurgical Unit. Both generators feature computer-controlled Instant Response Technology designed to ensure a consistent clinical effect through all tissue types. You can use handset or footswitch controls to activate the generator. Receptacles are marked to ensure the correct connection of accessories. Indicators alert you when the generator is activated. The Force FX™ allows you to adjust the activation tone volume and reset the mode and power settings used previously. The Force FX™ can be used in conjunction with the Valleylab Force GSU System and the Valleylab Electroshield Monitor. Refer to the manuals provided with these products for more information.
Description
ValleyLab Force FX and FXc Electrosurgical Units Features
The ValleyLab Force FX and FXc Electrosurgical Units are High Power Digital Electrosurgical Units. Both generators feature computer-controlled Instant Response Technology designed to ensure a consistent clinical effect through all tissue types. You can use handset or footswitch controls to activate the generator. Receptacles are marked to ensure the correct connection of accessories. Indicators alert you when the generator is activated. The Force FX™ allows you to adjust the activation tone volume and reset the mode and power settings used previously. The Force FX™ can be used in conjunction with the Valleylab Force GSU System and the Valleylab Electroshield Monitor. Refer to the manuals provided with these products for more information.
ValleyLab Force FX
ValleyLab Force FX and FXc Electrosurgical Units generators with computer-controlled Instant Response™ Technology are designed to ensure a consistent clinical effect through all tissue types.
ValleyLab Force FXc
- Instant Response™ technology ensures that the power delivered remains virtually constant, regardless of the tissue type
- Improved performance at lower power settings minimizes the risk of tissue damage and neuromuscular stimulation
- Three internal microcontrollers reduce system reaction time and increase the system’s processing speed
- Spray coagulation voltage of no more than 9000 volts peak-to-peak output for broad, but superficial coagulation with limited capacitive coupling
- A Power Efficiency Rating (PER) of approximately 98 for accurate and consistent cut performance
ValleyLab Force FX Electrosurgical Unit Specifications
Dimensions
- Height: 4.375″
- Width: 14″
- Length: 17″
Weight
- 18 lbs
Input Power Requirements
- Operating Range: 170 – 264
- VAC Max Current: 3.5 A in cut and 2 A in coag
Output Waveforms
- Bipolar
- Precise, standard and Macros: 470 kHz sinusoid
Monopolar Cut
- Low: 390 kHz sinusoid. Similar to the pure cut mode except the maximum voltage is limited to a lower value
- Pure: 390 kHz sinusoid. Blend: 390 kHz bursts of sinusoid, recurring at 27 kHz intervals, 50% duty cycle envelope
Monopolar Coag
- Desiccate: 240 kHz sinusoid repeated at 39 kHz, 8% duty cycle
- Fulgurate: 390 kHz damped sinusoidal bursts with a repetition frequency of 57 kHz into 300Ohm
- Spray: 390 kHz damped sinusoid bursts with randomized repetition centered at 28 kHz
- Frequencies include 21 kHz < f < 35kHz
- The output is further modulated by a random 250 Hz envelope with a variable duty cycle
Low-Frequency Leakage
- 50 – 60 Hz. Source current, patient leads, all outputs tied together
- Normal polarity, intact chassis ground < 10μA
- Normal polarity, ground open < 50μA. Reverse polarity, ground open < 50μA
- Sink current at high line, all points < 50μA
High Frequency Leakage
- Bipolar: < 59.2 mArms
- Monopolar: < 150 mArms
- CEM™: < 150 mArms at < or equal 50 W
- Power Readouts:
- Agree with actual power into the rated load to within ±15% or 5 watts, whichever is greater.