Philips Avalon FM30 Fetal Monitor

  • Philips Avalon FM30 Fetal Monitor

The Avalon FM30 Fetal Monitor offers an extensive set of external and internal fetal parameters, such as ultrasound, direct fetal heart rate, toco, and intrauterine pressure, as well as maternal pulse rate, and ECG, with optional SpO2 and blood pressure. View maternal and fetal ECG waves on the large, colorful display.

The Avalon FM30 fetal monitor also offers an extensive set of external and internal fetal parameters, such as ultrasound, direct fetal heart rate, toco, and intrauterine pressure, as well as maternal pulse rate, and ECG, with optional SpO2 and blood pressure. View maternal and fetal ECG waves on the large, colorful display.

Product Overview Button

Description

Philips Avalon FM30 Fetal Monitor Features

The Philips Avalon FM30 Fetal Monitor offers an extensive set of external and internal fetal parameters, such as ultrasound, direct fetal heart rate, toco, and intrauterine pressure, as well as maternal pulse rate, and ECG, with optional SpO2 and blood pressure. View maternal and fetal ECG waves on the large, colorful display. The Avalon FM30 fetal monitor also offers an extensive set of external and internal fetal parameters, such as ultrasound, direct fetal heart rate, toco, and intrauterine pressure, as well as maternal pulse rate, and ECG, with optional SpO2 and blood pressure. View maternal and fetal ECG waves on the large, colorful display.

Not only can the Avalon FM20 and FM30 monitor triplets, these advanced and compact fetal monitors are also simply smart enough to provide all the information you need to care for your patients:

  • Crisp and clear intuitive color touchscreen with large numerics and graphics
  • Backup memory for seamless monitoring
  • LAN interface for compatibility with hospital IT networks and access to additional system capabilities
  • Smart transducers with auto-recognition simplify operation
  • Transducer plug and play with automatic screen layout to focus on the patient, not the system
  • Data buffer with multi-patient handling for random printout
  • Cross-channel verification of maternal and fetal heart rates

Philips Avalon FM30 Fetal Monitor Specifications

Dimensions and Weight

  • M2702A/M2703A
    • Height: 5.2” (133 mm)
    • Width: 11.3” (286 mm)
    • Depth: 13.2” (335 mm)
    • Weight: 11.2 lbs (5.1 kg)
  • M2704A/M2705A
    • Height: 6.8 in (172 mm)
    • Width: 16.5 in (420 mm)
    • Depth: 14.6 in (370 mm)
    • Weight: 19.8 lbs (<9.0 kg)

Display

  • Monitor Screen: 6.5 in / 16.51 cm
  • Touchscreen Operation: Optional

Battery

  • Operating Time: up to 4 hours

Monitor Operating Environmental Conditions

  • Temperature Range: 0°C to 45°C (32°F to 113°F)
  • Humidity Rage: <95% relative humidity @ 40°C/104°F
  • Altitude Range: -500 to 3000 m/-1640 to 9840 ft.

Monitor Storage Environmental Conditions

  • Temperature Range: -20°C to 60°C (-4°F to 140°F)
  • Humidity Range: <90% relative humidity @ 60°C/140°F
  • Altitude Range: -500 to 13100 m/-1640 to 43000 ft.

Transducer Operating Environmental Conditions

  • Temperature Rage: 0°C to 40°C (32°F to 104°F)
  • Humidity Range: <95% relative humidity @ 40°C/104°F
  • Altitude Range: -500 to 3000 m/-1640 to 9840 ft.

Transducer Storage Environmental Contditions

  • Temperature Range: -20°C to 60°C (-4°F to 140°F)
  • Humidity Range: <90% relative humidity @ 60°C/140°F
  • Altitude Range: -500 to 13100 m/-1640 to 43000 ft.

SpO2 Environmental Conditions

  • Operating Temperature Range 0°C to 37°C (32°F to 98.6°F)

Fetal / Maternal Ultrasound

  • Measurement Method: Ultrasound Pulsed Doppler
  • Measurement Range: US 50 to 240 bpm
  • Display Resolution: 1 bpm
  • Printer Resolution: 1/4 bpm
  • Jitter @ 200 bpm: ≤ 3 bpm
  • Display Update Rate: 1/ Second
  • Average Output Power: P = (4.3 ± 0.4) mW
  • Peak-negative Acoustic Pressure: p_ = (33.9 ± 3.6) kPa
  • Output Beam Intensity (Iob): (= spatial average – temporal average intensity) Isata = (2.38 ± 0.75) mW/cm
  • Spatial-peak Temporal Average Intensity: I spta = (10.3 ± 2.2) mW/cm2
  • Effective Radiating Area: @ -6 dB 1.81 cm2
  • Beat to Beat Change: (max.) for Ultrasound 28 bpm
  • Frequency: 1 MHz ± 100 Hz
  • Signal Range: 3.5 μVpp to 350 μVpp @ 200 Hz

ECG

  • DECG: Single Lead ECG (derived from Fetal Scalp Electrode)
  • MECG: Single Lead ECG (derived from RA and LA electrodes)
  • Measurement Range: 30 to 240 bpm
  • Display Resolution: 1 bpm
  • Recorder Resolution: 1/4 bpm
  • Accuracy: ± 1 bpm or 1%, whichever is greater
  • Beat to Beat Change: (max.) 28 bpm
  • Differential Input Impedance: > 15MΩ
  • Electrode Offset Potential Tolerance: ± 400 mV
  • Filter Bandwidth: 0.8 to 80 Hz
  • Inop Auxiliary Current (Leads Off Detection): < 100 μA
  • Input Signal Range DECG: 20 μVpp to 6 mVpp
  • 20 μVpp to 6 mVpp MECG: 150 μVpp to 6 mVpp
  • Dielectric Strength: 1500 Vrms
  • Defibrillator Protection: None
  • ESU Protection: None

Non-invasive blood pressure

  • Measurement Ranges:
    • Systolic 30 to 270 mmHg (4 to 36 kPa)
    • Diastolic 10 to 245 mmHg (1.5 to 32 kPa)
    • Mean 20 to 255 mmHg (2.5 to 34 kPa)
  • Accuracy:
    • Max. Std. Deviation: 8 mmHg (1.1 kPa)
    • Max. Mean Error: ±5 mmHg (±0.7 kPa)
  • Pulse Rate Range Accuracy:
    • 40 to 300 bpm
    • 40 to 100 bpm: ±5 bpm
    • 101 to 200 bpm: ±5% of reading
    • 201 to 300 bpm: ±10% of reading
  • Measurement Time:
    • Typical at HR > 60bpm
    • Auto/manual: 30 seconds (adult)
    • Maximum time: 180 seconds (adult)
  • Cuff Inflation Time Typical for normal adult cuff: Less than 10 seconds
  • Initial Cuff Inflation Pressure: 165 ±15 mmHg
  • Auto Mode Repetition Times: 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45, 60 or 120 minutes

Parameters

  • External fetal parameters: US/Toco
  • Twin capability: Standard
  • Triplets capability: Optional
  • Internal fetal parameters: N/A
  • Maternal parameters: MECG, NIBP, Pulse
  • Smart Pulse technology: Standard
  • Cross channel verification: Standard
  • Fetal movement profile: Standard

Interfaces

  • OPTIONAL: PS/2 Interfaces
  • OPTIONAL: System Interface: Serial, LAN

Readings

  • OPTIONAL: NST trace interpretation: Up to three Fetal Heart Rates (FHR)
  • NST timer Standard

Data Storage

  • Data buffer Up to one hr