Criticare nGenuity Patient Monitor

  • Criticare nGenuity Patient Monitor

Improve patient outcomes with a comprehensive, user-configurable patient monitoring system covering the needs of multiple departments within hospitals and outpatient facilities.
The nGenuity™ patient monitor keeps you focused on your patients, not the monitor. When you demand accurate and reliable performance, The nGenuity™ patient monitor is a perfect choice.
The nGenuity™ patient monitor can monitor adult and, pediatric patients in virtually every department. For timely setups, the innovative user interface features simplified menus and, dedicated function keys are fast and easy to use. You can easily see the large numerics from across the room thanks to the big bright display.

Description

Criticare nGenuity Patient Monitor Features

The Criticare nGenuity Patient Monitor improves patient outcomes with a comprehensive, user-configurable patient monitoring system covering the needs of multiple departments within hospitals and outpatient facilities.

  • STANDARD: ETCO2, INCO2, ECG, ComfortCuff™ NIBP, DOX™ SpO2, HR, Temperature, Respiration, and Nurse Call interface connection.
  • Monitors End-tidal CO2 and Inspired CO2 accurately at high respiration rates.
  • Easy-to-read 10.4 active color TFT display.
  • Easy to learn, easy to use simplified menus and dedicated function keys.
  • Versatile mounting options.
  • OPTIONAL: Arrhythmia and ST analysis, printer.

Criticare nGenuity Patient Monitor Specifications

ECG

  • Number of Leads:
    • 3- or 5-Lead
  • Lead View:
    • User-selectable; I, II, III, aVR, aVL, aVF, V (5-Lead);
    • I, II, or III (3-Lead)
  • Gain Selection:
    • x0.5, x1, x2, and x4
  • Frequency Response:
    • Diagnostic: 0.05 to 100 Hz (-3dB);
    • Monitor: 0.5 to 40 Hz (-3dB);
    • ST: 0.05 to 40 Hz (-3dB)
  • Electrosurgery Protection:
    • Yes
  • Defibrillator Protection:
    • Yes
  • Pacer Detection/Rejection:
    • Yes
  • ST Accuracy:
    • ±0.2 mm or ±5% of reading, whichever is greater, over heart range of 30 to 120 bpm
  • ST Range:
    • -6.0 to 6.0 mm

Pulse Oximetry

  • Range:
    • 1% to 99%
  • Resolution:
    • 1%
  • Accuracy:
    • 70% to 99% range: ±2%
    • 50% to 69% range: ±3%
    • <50%: unspecified
  • Method:
    • Dual-wavelength

Non-Invasive Blood Pressure

  • Technique:
    • Oscillometric during inflation
    • Inflation Pressure Range: Adult: 30 to 300 mmHg
  • Pediatric:
    • 30 to 300 mmHg
  • Neonate:
    • 20 to 150 mmHg
  • Measurement Cycle:
    • < 40 sec. average
  • Automatic Measurement Cycles:
    • 2, 3, 5, 10, 15, 30, minutes
    • 1, 2, 4, hour
    • (User Selectable Cycle Times)
  • Stat Mode:
    • 5 minutes of continuous readings
  • Resolution:
    • 1 mmHg
  • Transducer Accuracy:
    • ±2 mmHg or ±2%, whichever is greater
  • STAT Mode:
    • 5 min of consecutive readings

Heart (Pulse) Rate

  • Source:
    • User selectable: Smart, ECG, Pleth, NIBP
  • Range:
    • 20 to 300 bpm (30-240 bpm, NIBP)
  • Accuracy:
    • ±1 bpm or 1% (ECG), whichever is greater
    • ±2 bpm or 2% (NIBP), whichever is greater

Alarms

  • Indication:
    • Audible; Visual
  • Silence:
    • Yes; 2 minutes or permanent

Temperature

  • Channels:
    • 1
  • Method:
    • Non-dispersive infrared; autocalibrating
  • Range, Accuracy:
    • 68° to 113° F (20° to 45° C): ±0.1° C
  • Display Resolution:
    • ±0.1° C
  • Resolution:
    • 1 mmHg; 0.1%; 0.1 kPa; 1 Torr
  • Probe:
    • YSI-400 and YSI-700 series

Trends

  • Memory:
    • 24 hours at 30-second intervals with an NIBP reading every 3 minutes
  • Display:
    • Tabular, Graphical

Display

  • Screen:
    • 10.4” active color TFT
  • Resolution:
    • 640 x 480 pixels
  • Waveforms:
    • 6, maximum
  • Waveform Types:
    • ECG Leads I, II, III, aVR, aVL, aVF, V; Respiration; CO2; Pleth; ST Histograms
  • Sweep Speeds:
    • 6.25, 12.5, 25, 50 mm/sec
  • Languages:
    • English, French, German, Italian, Portuguese, Spanish, Russian, and Chinese

Physical Properties

  • Weight:
    • 13.2 lb; 6 kg
  • Size:
    • 11” (H) x 13” (W) x 10.3” (D)
    • 28 cm (H) x 33 cm (W) x 26.2 cm (D)

Power Requirements

  • Voltage:
    • 100, 120, 220, 240 VAC; 50/60 Hz
  • Power Consumption:
    • 35 W, typical
  • Battery:
    • 1 sealed lead-acid battery; life: 3 hrs, typical;
  • Recharge time:
    • 4.5 hr

Environmental

  • Operating Temperature:
    • 59° to 95° F; 15° to 35° C
  • Storage Temperature:
    • 23° to 122° F; -5° to 50° C
  • Operating Humidity:
    • 15% to 90%, noncondensing

Classification

  • Medical Device:
    • Class II Equipment (IIb EU)
  • Electrical Protection:
    • Class I Equipment
  • Degree of Protection:
    • Type CF, Defibrillator-proof
  • Protection Against Ingress:
    • IPX1

Additional information

Brand

Criticare