Midmark Ritter M11

  • Midmark Ritter M11 Tabletop Autoclaves Sterilizers

The Midmark Ritter M11 is self-contained steam sterilizer that can be used in medical and dental offices, hospitals, clinics, nursing homes, laboratories, and other facilities to sterilize heat and moisture stable reusable items. The Midmark Ritter M11 is designed to optimize sterilization output, reducing time and effort.

Description

Midmark Ritter M11 Features

  • Intuitive display and simple prompts for easy operation
  • Four pre-programmed sterilization cycles: Unwrapped, Pouched, Packs, and Handpieces
  • Dry Cycle Door Stop – Prevents the door from opening full if there is a slight residual pressure in the chamber.
  • Monitors chamber temperature during a cycle to prevent an overheat condition.
  • Pressure relief valve opens to reduce chamber pressure if the chamber steam pressure exceeds the allowable limit.

Specifications

Midmark Ritter M11 Dimensions

  • Height (With Printer): 17.9 in. (45.2 cm)
  • Width: 17.8 in. (45.2 cm)
  • Depth (With Plug): 23.8 in. (60.5 cm)
  • Counter Area: 17.8 x 21 in (45.2 x 53.3 cm)
  • Chamber: 11.0 in diam. x 18 in deep (22.9 cm diam. x 46 cm deep)
  • Large Tray: 9 in x 15 in x 1 1/8 in (22.9 cm x 38 cm x 2.9 cm)
  • Small Tray: 6 5/8 in x 15 in x 1 1/8 in (14.3 cm x 38 cm x 2.9 cm)
  • Weight (Empty): 99 lbs (44.9 kg)
  • Weight (Full): 111 lbs (50.4 kg)

Water Reservoir Capacity

  • Approximately 1.4 Gallons (5.3 liters) to full mark. Usable volume is 1 Gallon (3.8 liters)

Electrical Requirements

  • 115 VAC, 50/60 Hz, 15 Amps Single Phase
  • 115 VAC has 1425 watt tubular immersion heater
  • 230 VAC, 50/60 Hz, 10 Amps Single Phase
  • 230 VAC has 1500 watt tubular immersion heater

Chamber Pressure

  • At 270 °F (132 °C) : 27.1 psi. (182 kPa)
  • Safety Valve Setting: 40 psi. (275.8 kPa)

Heat Emission

  • 5000 BTU/Hr during operation

Fuses

  • 115 VAC models:
    • F1: 0.250 amp, 250 V, Slo-Blo, 1/4” x 1-1/4”
    • F2: 15 amp, 250 V, Fast-Acting, 1/4” x 1-1/4”
  • 230 VAC models:
    • F1: 0.125 amp, 250 V, Slo-Blo, 5mm x 20mm
    • F2: 8 amp, 250 V, Fast-Acting, 5mm x 20mm

Safety Standards

  • UL Listed to U.S. & Canadian StandardsUL 61010-1, 2nd Edition
  • IEC 61010-2-040, 1st Edition
  • CAN/CSA C22.2 No. 61010-1, 2nd Edition
  • CSA C22.2 No. 61010-2-040-07, Part 2-040, 1st Edition
  • Meets the requirements of ASME Boiler and Pressure Vessel Code
  • A separate (dedicated) circuit is recommended for this sterilizer