Hospira Plum A+3 Infusion Pump

  • Hospira Plum A+3 Infusion Pump

The Hospira Plum A+3 Infusion Pump with Hospira MedNet Software is a single unit that allows clinicians to manage an industry-leading six-infusion line medication management system throughout the continuum of care. The wireless-enabled system assists healthcare professionals to better manage the complex medication dosing regimens typically seen in emergency rooms (ERs), intensive care units (ICUs), and oncology centers. Patients in these settings are often critically ill and may require multiple drug infusions delivered at precise doses and on varying schedules. The Plum A+3 with Hospira MedNet Software enables clinicians to deliver these infusions either simultaneously or in specific tiered sequences, depending on each patient’s needs.

Product Overview Button

Description

Hospira Plum A+3 Infusion Pump Features

The Hospira Plum A+3 Infusion Pump with Hospira MedNet Software is a single unit that allows clinicians to manage an industry-leading six-infusion line medication management system throughout the continuum of care. The wireless-enabled system assists healthcare professionals to better manage the complex medication dosing regimens typically seen emergency rooms (ERs), intensive care units (ICUs) and oncology centers. Patients in these settings are often critically ill and may require multiple drug infusions delivered at precise doses and on varying schedules. The Plum A+3 with Hospira MedNet Software enables clinicians to deliver these infusions either simultaneously or in specific tiered sequences, depending on each patient’s needs.

  • Deliver infusions with confidence with the unique PlumSet™ technology
  • Automated second-line delivery — eliminates the need to raise and lower infusion containers, improving ease-of-use and efficiency
  • Concurrent delivery of 2 medications through a single patient line
  • Air-in-Line detection and elimination — air trap captures up to 2 mLs of air
  • Accumulated air can be removed by automated back-priming without disconnecting the patient line, reducing the risk of contamination
  • Deliver secondary infusions via IV container or syringe
  • Reliable performance, steady, consistent and proven
  • One integrated system works across the entire spectrum of clinical care
  • Single channel (2 concurrent medications) or triple channel (6 concurrent medications) models available

Choose From a Variety of Programming Options:

  • Automated piggyback and concurrent delivery
  • Programmable standby settings
  • Multistep delivery
  • Loading dose automation
  • Programmable delayed starts

Hospira Plum A+3 Infusion Pump Specifications

Dimensions and Weight

  • Hight: 19 in
  • width: 15 in
  • Depth: 14 in
  • weight: 28 lbs (With 3 Batteries)

Electrical

  • Power Requirement: 120 V~, 50 – 60 HZ, 120 VA
  • Fuses: F1, F2, 250~, 0.5 A ( internal)
  • Power Cord: Hospital-grade AC cord. 10 ft long, with transparent plug and retainer plate
  • Battery: Three sealed, lead-acid, rechargeable 6 V batteries, internal to device.
  • Battery Life: With a new fully charged battery, the infuser shall operate for a minimum of six hours at 125 mL/hr or less, or deliver 500 mL at 126 mL/hr or greater on one line.
  • Recharge: The batteries charge whenever the pump(s) is connected to AC power. The recharge time is approximately six hours with the device operating at 125 mL/hr on one line.
  • Leakage:
    • Meets ANSI/AAMI ES1-1993 Standard
    • Safe Current Limits for Electro-medical Apparatus
  • Circuitry Ratings:
    • Voltage-30 VDC Max
    • Current- 0.25 Amps Max
    • Contact Rating- 3 Watts Max

Environment

  • Operating Temperature: 5° to 40° C
  • Storage Temperature: -20° to 60° C
  • Atmospheric Pressure: 0 – 10,000 feet
  • (0 – 3,000m) or equivalent pressure
  • Relative Humidity: 10 – 90% (40° C Max)

General

  • Volume Range: 0.1 to 99.9 mL (in 0.1 mL increments), 100 to 9999 mL (in mL increments)
  • Delivery Rate Accuracy: ± 5% (1-999 mL/hr), ± 10% (0.1-0.9 mL/hr)
  • Delivery Rates:
    • Primary: 0.1 mL/hr Minimum, 999 mL/hr Maximum
    • Secondary: 0.1 mL/hr Minimum, 999 mL/hr Maximum
    • Concurrent: 0.5 mL (per line) Minimum, 500 mL/hr (cumulative) Maximum
  • Delivery Units:
    • mL/hr
    • mcg/kg/min
    • mcg/min
    • mcg/kg/hr
    • cg/hr
    • mg/min
    • mg/kg/hr
    • mg/hr
    • ng/kg/min
    • g/hr
    • mEq/hr
    • Million units/hr
    • units/min
    • units/hr
    • units/kg/hr
    • units/kg/min
    • mmol/min
    • mmol/hr
  • System Alarms:
    • Air-in-Line
    • Air-in-Line Backpriming
    • Low Battery
    • Occlusion
    • Turn to Run
    • Flow Detector
    • VTBI Complete
    • Dose End
  • Occlusion Alarm Pressure: 1-15 psig (default 6 psig)
  • KVO Rate: 1.0 mL/hr or the last primary delivery rate, whichever is less

Hospira Infusion Pumps gives health care professionals unmatched safety, efficiency, and interoperability in one connected IV system.

When it comes to IV therapy, patient safety is your first concern. We can help. With a system that features the four-time Best in KLAS Plum 360™ and LifeCare PCA™ IV Pumps with ICU Medical MedNet™ safety software connecting you to more EHR vendors than anyone else, we can help you reduce medication errors, improve quality of care, streamline workflows, and maximize revenue capture.

Hospira also gives you unmatched data security backed by the industry’s first UL Cybersecurity Assurance Program certification. And our team of clinical and technical experts is here to provide you with customized configuration, implementation, and data analytics services to help you get the most out of your investment.