GE Dash 5000 Multiparameter Monitor

  • GE Dash 5000 Multiparameter Monitor Multiparameter and ECG Monitors

The GE Dash 5000 Multiparameter Monitor is a portable monitoring system that is flexible and easy to use. Multiple mounting solutions and a wide range of accessories ensure easy and flexible workplace configuration so caregivers can adapt to each patient's monitoring needs. The GE Dash 5000 monitor is designed for integration into commonly utilized existing wireless networks, helping you leverage your investment in your current IT infrastructure. Advanced clinical parameters including 12SL™ for simultaneous ECG monitoring, GE EK-Pro™ for multi-lead arrhythmia analysis, ICG, mainstream, and sidestream CO2 and GE DINAMAP® NIBP with Super STAT™. Networking options include wired and wireless Up to 7 waveforms and up to 4 invasive pressures High-resolution CRG trends for NICU monitoring.

 

Description

GE Dash 5000 Multiparameter Monitor Features

The GE Dash 5000 Multiparameter Monitor is a portable monitoring system that is flexible and easy to use.

  • Designed for optimized monitoring in all care applications
  • Advanced algorithms for all monitored parameters
  • Fully flexible alarm settings allow personalized monitoring
  • ECG acquisition through cables or wireless
  • Bright alarm light at the bedside helps to immediately locate the patient in critical situations
  • Multiple mounting solutions and a wide range of accessories ensure easy and flexible workplace configuration
  • 3 waveform screen views to maximize the size of parameter values and waveforms
  • Convenient care area specific Remote iTouch Pad enables operation on both sides of the patient’s bed and direct access to frequently used functions
  • Remote Screen through DashPort2 for an optimal central overview
  • CO2 module and extension Rack for additional parameters
  • A wide array of interfacing options with iConnect solutions
  • Both mainstream and low-flow sidestream CO2 measurement options for applications from neonatal to adult patients

GE Dash 5000 Multiparameter Monitor Specifications

Weight

  • 16 lb (7.3 kg)

Display

  • Screen Size: 12.1 in.
  • Type: Active-matrix color LCD.
  • Resolution: 800 by 600 dpi.
  • Number of traces: 7 (maximum).
  • Number of seconds/trace: 4.9 at 25 mm/sec.
  • Sweep speed: 6.25, 12.5, 25 mm/sec (with erase bar).

Controls

  • TrimKnob® Control
  • Five hard keys Standard Silence Alarm, Print, NBP Go/Stop, Zero All and Power On/Off.
  • Dash 5000 adds
    • Trends, NBP Auto, Admit/Discharge, Standby and Main View
    • Remote Control Option Available

Alarms

  • Categories Patient status and system status
  • Priority 4 levels – Crisis, Warning, Advisory and Message
  • Notification Audible and visual
  • Setting Default and individual
  • Silencing 1 minute, the current alarm only
  • Pause 5 minutes in Adult ICU mode, 3 minutes in Neonatal ICU mode and 5 minutes, 15 minute, or permanent pause in OR mode
  • Volume Default 70 dB measured at 1 meter

Invasive Blood Pressure

  • Number of channels 1 to 4 (optional)
  • Transducer sites Arterial, femoral artery, pulmonary arterial, central venous, right atrial, left atrial, intracranial and special
  • Transducer requirements Excitation voltage: 5 V dc ± 0.1%
  • Transducer output 5 μV/V/mmHg

Input Specifications

  • Range -25 mmHg to 300 mmHg
  • Offset ± 150 mmHg

Output Specifications

  • Frequency response dc to 50 Hz (-3 dB)
  • Zero balance range ± 150 mmHg
  • Zero balance accuracy ± 1 mmHg
  • Zero balance drift ± 1 mmHg over 24 hours
  • Accuracy ± 2% or ± 1 mmHg, whichever is greater (exclusive of the transducer)
  • Alarms User-selectable upper and lower limits for systolic, diastolic and mean pressures

ECG

  • Standard leads available I, II, III, V, aVR, aVL, and aVF
  • 3 lead wire I, II, or III
  • 5 lead wire I, II, III, V, aVR, aVL, and aVF
  • 10 lead wire I, II, III, aVR, aVL, aVF, VI, V2, V3, V4, V5 and V6
  • Leads analyzed simultaneously I, II, III, and V (multi-lead mode)
  • Lead Fail Identifies Failed Lead
  • Alarms User-selectable upper and lower heart rate limits

Input Specifications

  • Voltage range ± 0.5 mV to ± 5 mV
  • Signal width 40 ms to 120 ms (Q to S)
  • Heart rate range 30 to 300 bpm
  • Input impedance Common mode > 10 MΩ at 50/60 Hz
  • Differential > 2.5 MΩ from dc to 60 Hz
  • Common mode rejection 90 dB minimum at 50 or 60 Hz

Output Specifications

  • Impulse response For an impulse of 3 mV applied for 100 ms
  • Displacement following impulse < 0.1 mV
  • Slope following impulse < 0.3 mV/s
  • Frequency response Response of non-permanent displays is limited by resolution to 40 Hz (-3 dB) @25 mm/sec. Specified Upper Frequency Limits May Vary By +2 Hz.
  • Diagnostic mode 0.67 Hz (+0.4 dB) to 100 Hz (-3 dB)
  • For compliance with China 1.0 Hz (+0.4dB) to 75 Hz (-3 dB)
  • National Standard
  • Monitoring mode 0.67 (+0.4 dB) to 40 Hz (-3 dB)
  • Moderate mode 0.67 (+0.4 dB) to 25 Hz (-3 dB)
  • Maximum mode 5.0 Hz (-0.3 dB) to 25 Hz (-3 dB)
  • Noise < 30 μV (referred to input)

Pacemaker Detection / Rejection

  • Input voltage range ± 2 mV to ± 700 mV
  • Input pulse width 0.1 ms to 2 ms
  • Rise time 10 μs to 100 μs
  • Over/under shoot 2 mV (max)
  • Baseline drift < 0.5 mV per hour with a ± 700 mV, 2 ms
  • Pacemaker Pulse Applied

Respiration

  • Measurement technique Impedance variation detection
  • Range 0-200 breaths per minute for variations of 1.0 – 10.0 Ω
  • Respiration rate 0-200 breaths per minute
  • Base impedance 100-1000 Ω at 52.6 kHz
  • Detection sensitivity 0.4 to 10 Ω variation
  • Waveform display bandwidth 0.1 to 1.8 Hz (-3 dB)
  • Alarms User-selectable upper and lower respiration rate limits, and user-selectable apnea limit

Temperature

  • Number of channels: 2
  • Input Specifications
    • Probe type YSI Series 400 or 700 (determined by input cable)
    • Temperature range 0°C to 45°C (32°F to 113°F)
    • Resolution ± 0.1°C
  • Output Specifications
    • Parameters displayed T1, T2
    • Accuracy (independent of source) ± 0.1°C for YSI series 400; ± 0.3°C for YSI series 700 probes
    • Alarms User-selectable upper and lower limits for T1, T2

CARDIAC OUTPUT

  • Input Specifications
    • Probe type In-line or bath probe
    • Catheter size 5F, 6F, 7F, 7.5F and 8F
    • Injectate volume 3, 5 or 10 cc
  • Output Specifications
    • Parameters displayed Cardiac output, blood temperature, injectate temperature, and trial number
  • Range
    • Cardiac output 0.2-15 (liters per minute)
    • Blood temperature 30-42°C
    • Injectate temperature 0-30°C
  • Accuracy
    • Cardiac output ± 5%
    • Blood temperature ± 0.2°C
    • Injectate temperature ± 0.3°C
    • Frequency response dc to 15 Hz ± 2 Hz

Pulse oximetry

  • Parameters monitored Arterial oxygen saturation (SpO2) and peripheral pulse rate (PPR)
  • SpO2 range Nellcor 1-100%; Masimo 30-100%; GE Ohmeda 30-100%
  • PPR range Nellcor 20-300 BPM; Masimo 25-240 BPM; GE Ohmeda 30-250
  • Accuracy Actual accuracy depends on the probe. Please reference manufacturers’ specifications.
  • Nellcor SpO2 ± 2 digits (70-100% SpO2)
  • Masimo SpO2 ± 2% Adults/Pediatric (70-100% SpO2)
  • GE Ohmeda SpO2 ± 2% (70-100% SpO2), SpO2 ± 3% Neonates, ≤ 69% unspecified.
  • PPR ± 3 beats per minute
  • Alarms User-selectable upper and lower limits for SpO2 and PPR

DINAMAP Non-Invasive Blood Pressure

  • Technology DINAMAP® classic and SuperSTAT™ (SuperSTAT only available with Masimo and Nellcor SpO2)
  • Measurement technique Oscillometric
  • Displayed parameters Systolic, diastolic and mean pressures, time of last measurement
  • Measurement modes Adult ICU and OR modes; manual, auto and stat, neonatal mode; manual and auto
  • Systolic
    • Adult 30-285 mmHg
    • Pediatric 30-235 mmHg
    • Neonate 30-140 mmHg
  • Map
    • Adult 20-260 mmHg
    • Pediatric 20-220 mmHg
    • Neonate 20-125 mmHg
  • Diastolic
    • Adult 10-220 mmHg
    • Pediatric 10-210 mmHg
    • Neonate 10-110 mmHg

Pulse Rate, as Displayed in Tabular Trends or 3 Waveform Display

  • Adult 30-200 bpm
  • Pediatric 30-200 bpm
  • Neonate 30-200 bpm

Miscellaneous Specifications

  • Overall system accuracy Meets or exceeds SP 10-1992 AAMI standards
  • Automatic cycle times 0-4 hours
  • Tubing length 12 feet adult, 8 feet neonatal
  • Automatic Cuff Deflation Cycle Time Exceeding 3 Minutes (90 Seconds Neonatal), French mode – Cycle time exceeding 2 minutes (60 seconds neonatal), power off, or cuff pressure exceeds 294 mmHg (± 6 mmHg) adult, 250 (± 5 mmHg) pediatric, 147 (± 3 mmHg) neonatal
  • Cuff sizes Thigh, large adult, adult, small adult, child, infant and neonatal, sizes #5 – #1 and assorted long sizes
  • Alarms User-selectable upper and lower limits for systolic, diastolic and mean pressures