ERBE ICC 350 Electrosurgical Unit

The top-of-the-line ERBE ICC 350 Electrosurgical Unit heads ERBE's family of units for HF surgery. They've given it one-of-a-kind functionality for all-around use, with the features and abilities developed in close teamwork with leading surgeons from many different medical disciplines.

The ICC 350 has received worldwide attention and recognition; it is a step forward from conventional surgical methods and opens the way for new ones.

The ERBE ICC 350 Electrosurgical Unit is the best model for monopolar and bipolar cutting and coagulation. It's the universal, high-performance unit with a wide range of functions for every area application. Easy to use, with easy key-touch activation of complete settings for the whole unit (11 individual programs can be stored at the same time).

Description

ERBE ICC 350 Electrosurgical Unit Features

The ERBE ICC 350 Electrosurgical Unit is the best model for monopolar and bipolar cutting and coagulation.

  • General surgery
  • Dermatology
  • Endoscopy
  • Gastroenterology
  • Vascular Surgery
  • Gynecology
  • Neurosurgery
  • Plastic Surgery
  • Oral surgery

ERBE ICC 350 Electrosurgical Unit Specifications

Dimensions

  • Dimensions (W x H x D): 410 × 152 × 368 mm
  • Weight: 10.0 kg

Electrical

  • Rated power voltage: 240V/230V/115V/110V/100V± 10%
  • Rated power frequency: 50/60 Hz
  • Power current: 4.0 A at 230 – 240 V/8.0 A at 100-115 V
  • Power consumption in Standby mode: 25 watts
  • Power consumption at max. HF power: 620 watts 920 VA
  • Current consumption in Standby mode: 150 mA at 230 – 240V/300 mA at 100-115 V
  • Potential equalization terminal: Yes
  • Power fuses: 2 slow burn, 4 A at 230 – 240 V/8 A at 100-115 V

Cutting

  • HF power adjustment: With up/down increment keys
  • The form of HF voltage: Sine-shaped
  • The nominal frequency of HF voltage: 330 kHz
  • Cutting quality: 4 coagulation effects, selectable by key
  • The constancy of the coagulation effects: Automatically regulated
  • Support for initial incision and cutting: PPS, PowerPeakSystem
  • Monopolar AUTO CUT: With automatic voltage control
  • Monopolar HIGH CUT: with automatic arc control
  • Nominal HF output: 300 watts at 500 ohms
  • Peak HF power with PPS: 400 watts at 500 ohms
  • Monopolar ENDO CUT: 200 watts at max 200 ohms
  • BIPOLAR AUTO CUT: With automatic voltage control
  • Nominal HF output: 100 watts at 500 ohms

Coagulation

  • HF power adjustment: With up/down increment keys

Monopolar SOFT COAG

  • Form of HF current: Sine-shaped
  • Nominal frequency of HF voltage: 300 kHz
  • Nominal HF output: 120 watts at 125 ohms
  • End of coagulation: with Pedal or AUTO STOP

Monopolar FORCED COAG

  • Form of HF current: Pulse-modulated
  • Nominal frequency of HF voltage: 1 MHz
  • Nominal HF output: 120 atts at 350 ohms

Monopolar SPRAY COAG

  • Form of HF current: Pulse-modulated
  • Nominal frequency of HF voltage: 1 MHz
  • Nominal HF output: 120 watts at 500 ohms

Biopolar COAG

  • Form of HF current: Sine-shaped
  • The nominal frequency of HF voltage: 330 kHz
  • Nominal HF output: 120 watts at 125 ohms
  • Start of coagulation: By foot pedal or AUTOSTART
  • End of coagulation: By foot pedal or AUTO STOP

Safety Precautions

  • Safety class: I, corresponds to DIN IEC 601-1
  • Type: CF, corresponds to DIN IEC 601-1
  • Software check of LF leakage current, Icritical: 0.05mA (only with capacitive grounding)
  • Software check of HF leakage current, Icritical: 150 mA/300MA
  • NESSY: Neutral electrode Safety System
  • Neutral electrode switching: Switch-selectable capacitively grounded or with floating power
  • Self-Check: Includes all safety-related subassemblies

Additional information

Brand

ERBOTOM