Drager Perseus A500 Anesthesia Machine
Combine proven ventilation technology with the latest refinements in ergonomics and systems integration with an advanced anesthesia platform designed together with experts from around the world to streamline your anesthesia workflow.
Description
Drager Perseus A500 Anesthesia Machine Features
The Drager Perseus A500 Anesthesia Machine combines proven ventilation technology with the latest refinements in ergonomics and systems integration with an advanced, easy-to-use anesthesia platform designed together with experts from around the world to streamline your anesthesia workflow.
The Perseus A500 gives you a quality ventilation system that can support the patient’s breathing in all situations. Providing maneuvers that can automate various operating sequences. Adjustments can be made by using individual controls at any time. Because of the optimized breathing system, changes in the gas concentration reach the patient more readily, especially during low- and minimal-flow anesthesia.
- Ergonomic working conditions for users of different heights
- Workstation can be modified from standing into a sitting position within a few seconds
- Hygienic workplace thanks to the floor remaining clear and a bespoke cable management system
- Advanced ventilation in the operating room
- Enhanced workplace ergonomics
- Flexible workflow support
- Low- and minimal-flow anesthesia tools
- Award-winning design
Drager Perseus A500 Anesthesia Machine Specifications
Dimensions and Weight
- Height: 58.3 in (148 cm)
- Width: 45.2 in (115 cm)
- Depth: 31.1 in (79 cm)
- Weight: 335 lbs (150 kg)
Power
- Power Consumption 70 W, typical; max. 2.2 kW with aux power sockets in use
- Electrical mains connection (w/o isolation transformer): 100 – 240 V~ 50/60 Hz
- Electrical mains connection (w isolation transformer): 100 – 127 V~ 50/60 Hz or 200 – 240 V~ 50/60 Hz
- Maximum power consumption: 12 A
- Integrated battery backup time: 30 minutes minimum, 150 minutes typical (new and fully charged battery)
General
- Data interfaces: 2 x RS 232 (MEDIBUS protocol), 1 x USB, 1 x LAN
- Integrated power sockets: 4 x country specific (w isolation transformer) or 4 x IEC
- Latex: Contains no latex-containing components
- Storage surfaces and drawers:
- 3 drawers:
- 1 standard (lockable)
- 2 OPTIONAL (1 lockable)
- Ambient Conditions – Temperature: 10 to 40 °C (50 to 104 °F)
- Air Pressure: 620 to 1060 hPa (9.0 to 15.3 psi); Equivalent to 4.000 meter height
Fresh-Gas Delivery Electronic Mixer
- Fresh-Gas Flow: Off; 0.2 to 15 l/min
- O2-Concentration: 21 to 100 % in Air, 25 to 100 % (in N2O)
- O2-Flush: 25 to 75 l/min at 2.7 to 6.9 bar gas supply pressure
- O2-Flow for Auxiliary and Additional Oxygen: Off; 2 to10 l/min
Standard Modes of Ventilation
- Manual/Spontaneous (MAN/SPON)
- Pressure controlled: time-cycled (PC-CMV), synchronized (PC-BIPAP)
- Volume controlled: time-cycled (VC-CMV), time-cycled AutoFlow (VC-CMV/AF), synchronized AutoFlow (VC-SIMV/AF)
Optional Modes of Ventilation
- Pressure Support (CPAP/Pressure Support), selectable (PC-BIPAP/PS) and Autoflow Pressure support for pressure-controlled ventilation (VC-SIMV/AF/PS), selectable CPAP for Man/Spont
- Airway Pressure Release Ventilation (PC-APRV)
- External fresh gas outlet (OPTIONAL)
Other
- Patient Demographics: Neonates, pediatric patients, adults
- Tidal Volume: 20 to 2,000 ml
- Inspiratory Pressure PINSP: 3 to 80 hPa/mbar/cmH2O
- Pressure Limitation PMAX: 7 to 80 hPa/mbar/cmH2O
- Pressure Support Over PEEP: 0 to 78 hPa/mbar/cmH2O
- Respiratory Rate: 3 to 100/min
- Inspiratory Time: 0.2 to 10 s
- Inspiratory Flow: 0 to 180 l/min
- PEEP/CPAP: Off, 2 to 35 hPa/mbar/cmH2O
Additional information