Drager Infinity Gamma XL ECG Multiparameter Monitor

  • Drager Infinity Gama XL ECG Multiparameter Monitor

The Drager Infinity Gamma XL ECG Multiparameter Monitor configured patient monitors, now with VF5 software functionality, provide a set of the most commonly used parameters in a compact, user-friendly package. Because of the unique Pick and Go‚® capabilities, these monitors show continuous surveillance while eliminating the need to purchase separate transport monitors. Patient data from both monitors is available either hardwired or wireless over the Infinity Network.

Description

Drager Infinity Gamma XL ECG Multiparameter Monitor Features

The Drager Infinity Gamma XL ECG Multiparameter Monitor configured patient monitors, now with VF5 software functionality, provide a set of the most commonly used parameters in a compact, user-friendly package. Because of the unique Pick and Go‚® capabilities, these monitors show continuous surveillance while eliminating the need to purchase separate transport monitors. Patient data from both monitors is available either hardwired or wireless over the Infinity Network.

  • Provide monitoring versatility for any sub-acute care setting
  • Display 3 or 4 channels and offer a choice of sizes
  • Provide continuous monitoring and data collection

Drager Infinity Gamma XL ECG Multiparameter Monitor Specifications

Dimensions and Weight

  • Width: 11.9 in. (301 mm)
  • Height: 8.5 in. (215 mm)
  • Depth: 5.2 in. (131 mm)
  • Weight: 3.32 kg (7.32 lbs.) without battery

Monitoring Capabilities

  • Neonatal, pediatric and adult applications

ECG

  • Available leads
    • 3-lead set Leads: I, II, III
    • 5-lead set Leads: I, II, III, aVR, aVL, aVF, V
    • 6-lead set Leads: I, II, III, aVR, aVL, aVF, V, V+
  • Measuring Range: 15 to 300 bpm
  • Accuracy: ± 5 bpm or ± 2% (whichever is greater)
  • Frequency range
    • (± 3 db)
    • 0.5 Hz to 28 Hz (50 Hz)
    • 0.5 Hz to 40 Hz (60 Hz)
  • Filter Frequency: 50 to 60 Hz

QRS Detection Range

  • Alarms User-selectable upper and lower limits
  • Pacer detection Leads: I, II, or III
  • Amplitude: ± 5 to ± 700 mV
  • Separation between pulses: 30 msec
  • Width (dp): 0.1 to 2.0 msec
  • Accessories 3-, 5- or 6-lead electrode set

Dual Lead ST Analysis

  • Refer to ST datasheet 9049456

Respiration

  • Sensing lead II
  • Measuring method Impedance pneumography
  • Auxiliary current 10μA for any active electrode
  • Detection threshold 0.15 to 4.0 in manual mode (user adjustment)
  • Adult and pediatric: 0.20 to 10.5 in auto mode (automatic adjustment)
  • Neonatal: 0.20 to 1.5 in auto mode (automatic adjustment)
  • Measuring Range: 0 to 155 breaths/min
  • Accuracy: ± 1 l/min or ± 2% (whichever is greater)
  • Alarms User-selectable upper and lower respiration rate

Pulse Oximetry (SpO2)

  • Displayed parameters Saturation (fraction of oxyhemoglobin to functional hemoglobin) and pulse (rate and curve)
  • Measuring Method: Absorption Spectrophotometry
  • Pulse: 15 to 300 bpm
  • Measuring Range: 0 – 69%
  • SpO2: 70 to 100% ± 2% (except ± 3% when using Nellcor® DS-100A and ± 3.5% when using Masimo® LNOP-Ear)
  • Saturation accuracy increases by ± 1 digit for neonates
  • Pulse: ± 3 bpm or ± 3% (whichever is less)
  • Alarms User-selectable upper and lower limits for SpO2 and pulse
  • Accessories Dräger Medical-approved Masimo and Nellcor sensors

Temperature

  • Displayed parameters Absolute temperature
  • Measuring Range: 0 to 50°C
  • Accuracy Probe: ± 0.1°C
  • System: ± 0.2°C
  • Alarms User-selectable upper and lower limits
  • Accessories Dräger Medical-approved core and skin probes

Noninvasive Blood Pressure (NBP)

  • Displayed parameters Systolic, mean and diastolic pressures
  • Measuring Method: Oscillometric Utilizing Step Deflation

Modes of Operation

  • Manual (single measurement) interval or continuous (5 minutes)
  • Interval times Off, 2, 2.5, 3, 5 (increments of 5), 30, 45, 60, 120, 180, 240 minutes

Measuring range (default)

  • Adult (Adult 270 mmHg) (Adult 180 mmHg)
    • Heart rate 30 to 240 bpm 30 to 240 bpm
    • Systolic 30 to 250 mmHg 30 to 170 mmHg
    • Mean 20 to 230 mmHg 20 to 150 mmHg
    • Diastolic 10 to 210 mmHg 10 to 130 mmHg
  • Pediatric (Pediatric 180 mmHg) (Pediatric 140 mmHg)
    • Heart rate 30 to 240 bpm 30 to 240 bpm
    • Systolic 30 to 170 mmHg 30 to 130 mmHg
    • Mean 20 to 150 mmHg 20 to 110 mmHg
    • Diastolic 10 to 130 mmHg 10 to 130 mmHg
  • Neonatal (Neonatal 140 mmHg) (Neonatal 90 mmHg)
    • Heart rate 30 to 240 bpm 30 to 240 bpm
    • Systolic 30 to 130 mmHg 30 to 80 mmHg
    • Mean 20 to 110 mmHg 20 to 70 mmHg
    • Diastolic 10 to 100 mmHg 10 to 60 mmHg

Cuff pressure

  • Default inflation pressure Adult (270): 180 mmHg ± 5 mmHg
  • Pediatric (180): 130 mmHg ± 5 mmHg
  • Neonatal (140): 110 mmHg ± 5 mmHg
  • Adult (180): 130 mmHg ± 5 mmHg
  • Pediatric (140): 110 mmHg ± 5 mmHg
  • Neonatal (90): 80 mmHg ± 5 mmHg
  • Inflation pressure after a valid measurement (± 5 mmHg) Adult (270): Systolic + 25 mmHg
  • Pediatric (180): Systolic + 25 mmHg
  • Neonatal (140): Systolic + 30 mmHg
  • Adult (180): Systolic + 25 mmHg
  • Pediatric (140): Systolic + 30 mmHg
  • Neonatal (90): Systolic + 20 mmHg
  • Maximum inflation pressure Adult (270): 265 mmHg ± 5 mmHg
  • Pediatric (180): 180 mmHg ± 5 mmHg
  • Neonatal (140): 142 mmHg ± 5 mmHg
  • Adult (180): 180 mmHg ± 5 mmHg
  • Pediatric (140): 142 mmHg ± 5 mmHg
  • Neonatal (90): 80 mmHg ± 5 mmHg
  • Minimum inflation pressure Adult (270): 110 mmHg ± 5 mmHg
  • Adult (180): 90 mmHg ± 5 mmHg
  • Pediatric (180): 90 mmHg ± 5 mmHg
  • Pediatric (140): 70 mmHg ± 5 mmHg
  • Neonatal (both): 70 mmHg ± 5 mmHg
  • Connector Quick-release connector with single airway
  • Minimum Inflation Presure Adult (270): 110 mmHg ± 5 mmHg
  • Adult (180): 90 mmHg ± 5 mmHg
  • Pediatric (180): 90 mmHg ± 5 mmHg
  • Pediatric (140): 70 mmHg ± 5 mmHg
  • Neonatal (both): 70 mmHg ± 5 mmHg
  • Connector Quick-release connector with single airway

Invasive Blood Pressure

  • Displays up to 2 pressures (with IBP2 option)
  • Measuring method Resistive strain gauge transducer
  • Display resolution 1 mmHg
  • Measuring range -50 to 399 Hg
  • Frequency range DC to 16 Hz
  • Zero balance range ± 190 mmHg
  • Transducer specifications Drager medical approved transducers with a risistance of 300 – 2000 and an equivalent pressure sensitivity of 5μV/V/mmHg ± 10%
  • Accuracy ± 2 mmHg or ± 3%, after successful zero and calibration (exclusive of transducer)
  • IBP alarms User-selectable upper and lower limits for systolic, diastolic and mean pressures
  • Accessories Dräger Medical-approved pressure transducers

Display Specifications

  • Type Thin Film Transistor-Liquid Crystal Display
  • Active Matrix (TFT-LCD)

Size

  • Gamma XL 21 cm (8.4 in.) diagonal
  • Gamma 16.5 cm (6.5 in.) diagonal

Channels

  • Gamma XL 4 standard
  • Gamma 3 standard with 4 channel option

Viewing Area

  • Gamma XL 170.9 mm x 129.6 mm
  • Gamma 132.5 mm x 99.4 mm
  • Resolution 640 x 480 pixels

User Interface

  • Rotary Knob: For Selection and Confirmation of Choices
  • Easy-to-use Menu Structure and Fixed Keys
  • Alarm levels Life-threatening, serious and advisory

Information Management Capabilities

  • Data storage 24 hours
  • Data resolution 60 seconds
  • Trend tables 1, 5, 15, 30 or 60 minute display formats
  • Trend graphs 1, 2, 4, 8 or 12 hour display format
  • Trend type Tabular and graphical tables and event recall (10 events)

Electrical Specifications

  • Input voltage 11 to 14V DC, 2.5A
  • Power consumption 52 watts (fully loaded)
  • Patient leakage current 10μA
  • Protection class Internally powered (per IEC 60601-1)

Power Adapter

  • Power requirements 100 to 120V AC, 0.8A 200 to 240V AC, 0.4A
  • Frequency 50 to 60 Hz
  • Chassis leakage current 300μA @ 110V AC and 60 Hz 500μA @ 220V AC and 50 Hz

Battery Specifications

  • Internal battery
    • Battery type Lead-acid
    • Battery capacity 75 minutes
    • Charging time 5.5 hours at 25°C
  • External auxiliary battery
    • Battery type Lithium-ion
    • Battery capacity 180 minutes
    • Charging time 8 hours at 25°C
    • Battery life is specified under the following conditions:
      • NBP measurement every 15 minutes connected, no etCO2 and running at 25°C
      • Battery life may diminish after extended use.

Environmental Requirements

  • Temperature range
    • Operating 0° C to 45° C (without recorder)
    • Storage -20° C to 50° C
  • Relative humidity
    • Operating >10% and <95%, non-condensing
    • Storage >10% and <95% non-condensing
  • Atmospheric pressure
    • Operating 525 to 795 mmHg (70 to 106 kPa)
    • Storage 375 to 795 mmHg (50 to 106 kPa)