Datex Ohmeda Aestiva 5 Anesthesia Machine
The Datex Ohmeda Aestiva 5 Anesthesia Machine offers a good combination of quality and performance at an affordable price. The 7900 ventilators offer a full range of operational parameters including optimal tidal volume compensation, integrated volume, and oxygen monitoring. The machine also features Tek 5 Vaporizers. The Aestiva 5 has been optimized for patient safety through patient-centered engineering while also being ergonomically configured for the clinician’s comfort.
Description
Datex Ohmeda Aestiva 5 Anesthesia Machine Features
The Datex Ohmeda Aestiva 5 Anesthesia Machine offers a good combination of quality and performance at an affordable price.
- The Aestiva 5 offers a unique combination of quality performance and cost-control.
- The 7100 ventilator provides a wide range of capabilities, including optimal tidal volume compensation and integrated volume and oxygen monitoring.
- Recognized as an anesthesia system that has patient centered engineering and clinician-oriented efficiency
- Support for up to 8 waveforms on screen
- Anesthesia dedicated alarm system
- Alarms display and adjustment page
- Automatic alarm limits function
- Optional: Tek 5 Vaporizers
- Four display options with intergrated monitor controls:
- 15″ or 17″ High Resolution Color Video Display
- 10.4” or 15” Flat Panel Color LCD Display or Large 17” Color CRT Display
- User Configurable Display Layout and Colors
- Support for User Configurable Second and Third Screen with Different Waveform and Numarical Information
- 10-minute graphical mini trends referenced to set alarm limits
- Six User Configurable Modes and Special Cardiac Bypass Mode
- 24 hour numerical and graphical trends
- Split screen capability to show ST, spirometry loop, EEG, EP or mini trends permanently
- Four User Configurable Graphical Trend Pages
- Trend cursor in both graphical and numerical trends
- Waveform snapshots to store and recall waveform samples
- Help function
- On-line help for menu functions
- Open software and hardware architecture
- Easy insertion/removal of plug-in parameter modules without interrupting other monitoring
- Anesthesia care oriented user interface
- Optional extended arrhythmia analysis
- Connector for Datex-Ohmeda S/5™ Device Interfacing Solution
- Anesthesia Comand Bar for Intuitive Access to the Monitoring Functions
- Removable command bar brings monitor controls close to the user
- Frame with additional module slots
- Moves patient cables close to the patient
- Single cable connects the Extension Frame to the s/5 anesthesia Monitor’s Central Unit
- This will Reduce Patient Cable Clutter
- Mounts to any vertical or horizontal pole
Datex Ohmeda Aestiva 5 Anesthesia Machine Specifications
Dimensions | 2 Vaporizer Configuration | 3 Vaporizer Configuration |
---|---|---|
Height | 53.4″ (135.8 cm) | 53.4″ (135.8 cm) |
Width | 29.5″ (75 cm) | 36.6″ (93 cm) |
Depth | 32.7″ (83 cm) | 32.7″ (83 cm) |
Weight | 300 lb (136 kg) | 340 lb (154 kg) |
Surface Dimensions | Work Surface | Top Drawer (1 Standard) | Folding Side Shelf (Optional) | Lower Drawers (Optional) |
---|---|---|---|---|
Height | 34.5″ (87.5 cm) | 4.1″ (10.5 cm) | 34.5″ (87.5 cm) | 5.7″ (14.5 cm) |
Width | 18.5″ (47 cm) | 15.2″ (38.5 cm) | 10.4″ (26.5 cm) | 15.2″ (38.5 cm) |
Depth | 12.4″ (31.5 cm) | 10.2″ (26 cm) | 12.4″ (31.5 cm) | 10.2″ (26 cm) |
Weight Limit | 25 lb (11.3 kg) |
Top Shelves (Optional) | 2 Vaporizer Configuration | 3 Vaporizer Configuration |
---|---|---|
Width | 18.7, 26.6 or 34.4″ (47.5, 67.5 or 87.5 cm) | 34.4 or 26.6″ (87.5 or 67.5 cm) |
Depth | 16.1″ (41 cm) | 16.1″ (41 cm) |
Weight Limit | 100 lb (46 kg) | 100 lb (46 kg) |
Absorber Bag Arms | Adjustable | Non Adjustable |
---|---|---|
Arm Length | 12″ (30.5 cm) | 10″ (25.4 cm) |
Bag Arm Height | 34.3 or 40.9″ (87 or 104 cm) | 36″ (91.5 cm) |
Absorber Rotation | 85° | 85° |
Ventilator Screen
- Height: 3″ (7.6 cm)
- Width: 6″ (15.2 cm)
Casters
- Diameter: 5″ (12.5 cm)
- Brakes: Single foot lever locks and unlocks two front casters
DIN Rail (Optional)
- Side of Tabletop: 12″ (30 cm)
- Side of Machine: 9.25″ (23.5 cm)
Ventilator (VT) Parameter Ranges
- Tidal volume range:
- 20 to 1500 mL (Volume Control and SIMV modes)
- 5 to 1500 mL (Pressure Control Mode)
- Incremental settings:
- 20 to 100 mL (increments of 5 mL)
- 100 to 300 mL (increments of 10 mL)
- 300 to 1000 mL (increments of 25 mL)
- 1000 to 1500 mL (increments of 50 mL)
- Minute volume range:
- 0 to 99.9 L/min
- Pressure (PInspired) range:
- 5 to 60 cm H2O (increments of 1 cm H2O)
- Pressure (PLimit) range:
- 12 to 100 cm H2O (increments of 1 cm H2O)
- Pressure (PSupport) range:
- Off, 2 to 40 cm H2O (increments of 1 cm H2O)
- Rate:
- 4 to 100 breaths per minute for Volume Control and Pressure Control;
- 2 to 60 breaths per minute for SIMV, PSVPro and SIMV–PC+PSV (increments of 1 breath per minute)
- Inspiratory/expiratory ratio:
- 2:1 to 1:8 (increments of 0.5)
- Inspiratory time:
- 0.2 to 5.0 seconds (increments of 0.1 seconds) (SIMV and PSV Pro)
- Trigger window:
- 0 to 80% (increments of 5%)
- Flow trigger:
- 0.2 to 1.0 L/min (increments of 0.2 L/min) 1 to 10 L/min (increments of 0.5 L/min)
- Inspiration termination level:
- 5 to 75% (increments of 5%)
- Backup mode delay:
- 10 to 30 seconds (increments of 5 seconds)
Positive End Expiratory Pressure (PEEP)
- Type:
- Integrated, electronically controlled
- Range:
- OFF, 4 to 30 cm H2O (increments of 1 cm H2O)
Ventilator Performance
- Pressure range at inlet:
- 240 kPa to 700 kPa/35 psig to 100 psig
- Peak gas flow:
- 120 L/min + fresh gas flow
- Flow valve range:
- 1 to 120 L/min
- Flow compensation range:
- 200 mL/min to 15 L/min
Ventilator Monitoring
- Expiratory minute volume range:
- 0 to 99.9 L/min
- Expiratory tidal volume range:
- 0 to 1500 mL
- O2%:
- 5 to 110%
- Peak pressure:
- –20 to 120 cm H2O
- Mean pressure:
- –20 to 120 cm H2O
- Plateau pressure:
- 0 to 120 cm H2O
- Pressure waveform sweep speed:
- 4 to 25 breaths per minute (0 to 15 seconds)
- 26 to 75 breaths per minute (0 to 5 seconds)
- 75 breaths per minute (0 to 3 seconds)
Ventilator Accuracy
Delivery/Monitoring Accuracy
- Volume delivery:
- > 210 mL = better than 7%
- < 210 mL = better than 15 mL
- < 60 mL = better than 10 mL
- Pressure delivery:
- ±10% or ±3 cm H2O
- PEEP delivery:
- ±1.5 cm H2O
- Volume monitoring:
- > 210 mL = better than 9%
- < 210 mL = better than 18 mL
- < 60 mL = better than 10 mL
- Pressure monitoring:
- ±5% or ±2 cm H2O
Alarm Settings
- Tidal Volume (VTE):
- Low: OFF, 0 to 1500 mL
- High: 20 to 1600 mL, OFF
- Minute Volume (VE):
- Low: OFF, 0 to 10 L/min
- High: 0 to 30 L/min, OFF
- Inspired Oxygen (FlO2):
- Low: 18 to 100%
- High: 18 to 100%, OFF
- Apnea Alarm:
- Mechanical ventilation ON: > 5 mL breath measured in 30 seconds
- Mechanical ventilation OFF: > 5 mL breath measured in 30 seconds
- Low Airway Pressure:
- 4 cm H2O above PEEP
- High Pressure:
- 12 to 100 cm H2O (increments of 1 cm H2O)
- Sustained Airway Pressure:
- Mechanical Ventilation ON:
- PLimit < 30 cm H2O, the sustained limit is 6 cm H2O
- PLimit 30 to 60 cm H2O, the sustained limit is 20% of PLimit
- PLimit > 60 cm H2O, the sustained limit is 12 cm H2O
- PEEP and Mechanical Ventilation ON:
- Sustained limit increases by PEEP minus 2 cm H2O
- Mechanical ventilation OFF:
- PLimit 60 cm H2O, the sustained limit is 50% of PLimit
- PLimit > 60 cm H2O, the sustained limit is 30 cm H2O
- Mechanical Ventilation ON:
- Subatmospheric Pressure:
- Paw < -10 cm H2O
- Alarm Silence Countdown Timer:
- 120 to 0 seconds
Ventilator Components
Flow Transducer
- Type:
- Variable orifice flow sensor
- Dimensions:
- 22 mm OD and 15 mm ID
- Location:
- Inspiratory outlet and expiratory inlet
- Optional autoclavable sensor available
Oxygen Sensor
- Type:
- Galvanic fuel cell
- Life Cycle:
- Approximately 18 months (dependent on usage)
Anesthetic Agent Delivery
- Vaporizers:
- Tec 4, Tec 5, Tec 6 Plus, Tec 7
- Number of positions:
- 2 or 3
- Mounting:
- Tool-free installation Selectatec® manifold interlocks and isolates vaporizers
Additional information
Brand | GE |
---|