Datex Ohmeda Aestiva 5 Anesthesia Machine

  • Anesthesia Machines Datex Ohmeda Aestiva 5 Anesthesia Machine

The Datex Ohmeda Aestiva 5 Anesthesia Machine offers a good combination of quality and performance at an affordable price. The 7900 ventilators offer a full range of operational parameters including optimal tidal volume compensation, integrated volume, and oxygen monitoring. The machine also features Tek 5 Vaporizers. The Aestiva 5 has been optimized for patient safety through patient-centered engineering while also being ergonomically configured for the clinician’s comfort.

Description

Datex Ohmeda Aestiva 5 Anesthesia Machine Features

The Datex Ohmeda Aestiva 5 Anesthesia Machine offers a good combination of quality and performance at an affordable price.

  • The Aestiva 5 offers a unique combination of quality performance and cost-control.
  • The 7100 ventilator provides a wide range of capabilities, including optimal tidal volume compensation and integrated volume and oxygen monitoring.
  • Recognized as an anesthesia system that has patient centered engineering and clinician-oriented efficiency
  • Support for up to 8 waveforms on screen
  • Anesthesia dedicated alarm system
  • Alarms display and adjustment page
  • Automatic alarm limits function
  • Optional: Tek 5 Vaporizers
  • Four display options with intergrated monitor controls:
    • 15″ or 17″ High Resolution Color Video Display
    • 10.4” or 15” Flat Panel Color LCD Display or Large 17” Color CRT Display
  • User Configurable Display Layout and Colors
  • Support for User Configurable Second and Third Screen with Different Waveform and Numarical Information
  • 10-minute graphical mini trends referenced to set alarm limits
  • Six User Configurable Modes and Special Cardiac Bypass Mode
  • 24 hour numerical and graphical trends
  • Split screen capability to show ST, spirometry loop, EEG, EP or mini trends permanently
  • Four User Configurable Graphical Trend Pages
  • Trend cursor in both graphical and numerical trends
  • Waveform snapshots to store and recall waveform samples
  • Help function
  • On-line help for menu functions
  • Open software and hardware architecture
  • Easy insertion/removal of plug-in parameter modules without interrupting other monitoring
  • Anesthesia care oriented user interface
  • Optional extended arrhythmia analysis
  • Connector for Datex-Ohmeda S/5™ Device Interfacing Solution
  • Anesthesia Comand Bar for Intuitive Access to the Monitoring Functions
  • Removable command bar brings monitor controls close to the user
  • Frame with additional module slots
  • Moves patient cables close to the patient
  • Single cable connects the Extension Frame to the s/5 anesthesia Monitor’s Central Unit
    • This will Reduce Patient Cable Clutter
    • Mounts to any vertical or horizontal pole

Datex Ohmeda Aestiva 5 Anesthesia Machine Specifications

Dimensions 2 Vaporizer Configuration 3 Vaporizer Configuration
Height 53.4″ (135.8 cm) 53.4″ (135.8 cm)
Width 29.5″ (75 cm) 36.6″ (93 cm)
Depth 32.7″ (83 cm) 32.7″ (83 cm)
Weight 300 lb (136 kg) 340 lb (154 kg)
Surface Dimensions Work Surface Top Drawer (1 Standard) Folding Side Shelf (Optional) Lower Drawers (Optional)
Height 34.5″ (87.5 cm) 4.1″ (10.5 cm) 34.5″ (87.5 cm) 5.7″ (14.5 cm)
Width 18.5″ (47 cm) 15.2″ (38.5 cm) 10.4″ (26.5 cm) 15.2″ (38.5 cm)
Depth 12.4″ (31.5 cm) 10.2″ (26 cm) 12.4″ (31.5 cm) 10.2″ (26 cm)
Weight Limit 25 lb (11.3 kg)
Top Shelves (Optional) 2 Vaporizer Configuration 3 Vaporizer Configuration
Width 18.7, 26.6 or 34.4″ (47.5, 67.5 or 87.5 cm) 34.4 or 26.6″ (87.5 or 67.5 cm)
Depth 16.1″ (41 cm) 16.1″ (41 cm)
Weight Limit 100 lb (46 kg) 100 lb (46 kg)
Absorber Bag Arms Adjustable Non Adjustable
Arm Length 12″ (30.5 cm) 10″ (25.4 cm)
Bag Arm Height 34.3 or 40.9″ (87 or 104 cm) 36″ (91.5 cm)
Absorber Rotation 85° 85°

Ventilator Screen

  • Height: 3″ (7.6 cm)
  • Width: 6″ (15.2 cm)

Casters

  • Diameter: 5″ (12.5 cm)
  • Brakes: Single foot lever locks and unlocks two front casters

DIN Rail (Optional)

  • Side of Tabletop: 12″ (30 cm)
  • Side of Machine: 9.25″ (23.5 cm)

Ventilator (VT) Parameter Ranges

  • Tidal volume range:
    • 20 to 1500 mL (Volume Control and SIMV modes)
    • 5 to 1500 mL (Pressure Control Mode)
  • Incremental settings:
    • 20 to 100 mL (increments of 5 mL)
    • 100 to 300 mL (increments of 10 mL)
    • 300 to 1000 mL (increments of 25 mL)
    • 1000 to 1500 mL (increments of 50 mL)
  • Minute volume range:
    • 0 to 99.9 L/min
  • Pressure (PInspired) range:
    • 5 to 60 cm H2O (increments of 1 cm H2O)
  • Pressure (PLimit) range:
    • 12 to 100 cm H2O (increments of 1 cm H2O)
  • Pressure (PSupport) range:
    • Off, 2 to 40 cm H2O (increments of 1 cm H2O)
  • Rate:
    • 4 to 100 breaths per minute for Volume Control and Pressure Control;
    • 2 to 60 breaths per minute for SIMV, PSVPro and SIMV–PC+PSV (increments of 1 breath per minute)
  • Inspiratory/expiratory ratio:
    • 2:1 to 1:8 (increments of 0.5)
  • Inspiratory time:
    • 0.2 to 5.0 seconds (increments of 0.1 seconds) (SIMV and PSV Pro)
  • Trigger window:
    • 0 to 80% (increments of 5%)
  • Flow trigger:
    • 0.2 to 1.0 L/min (increments of 0.2 L/min) 1 to 10 L/min (increments of 0.5 L/min)
  • Inspiration termination level:
    • 5 to 75% (increments of 5%)
  • Backup mode delay:
    • 10 to 30 seconds (increments of 5 seconds)

Positive End Expiratory Pressure (PEEP)

  • Type:
    • Integrated, electronically controlled
  • Range:
    • OFF, 4 to 30 cm H2O (increments of 1 cm H2O)

Ventilator Performance

  • Pressure range at inlet:
    • 240 kPa to 700 kPa/35 psig to 100 psig
  • Peak gas flow:
    • 120 L/min + fresh gas flow
  • Flow valve range:
    • 1 to 120 L/min
  • Flow compensation range:
    • 200 mL/min to 15 L/min

Ventilator Monitoring

  • Expiratory minute volume range:
    • 0 to 99.9 L/min
  • Expiratory tidal volume range:
    • 0 to 1500 mL
  • O2%:
    • 5 to 110%
  • Peak pressure:
    • –20 to 120 cm H2O
  • Mean pressure:
    • –20 to 120 cm H2O
  • Plateau pressure:
    • 0 to 120 cm H2O
  • Pressure waveform sweep speed:
    • 4 to 25 breaths per minute (0 to 15 seconds)
    • 26 to 75 breaths per minute (0 to 5 seconds)
    • 75 breaths per minute (0 to 3 seconds)

Ventilator Accuracy

Delivery/Monitoring Accuracy

  • Volume delivery:
    • > 210 mL = better than 7%
    • < 210 mL = better than 15 mL
    • < 60 mL = better than 10 mL
  • Pressure delivery:
    • ±10% or ±3 cm H2O
  • PEEP delivery:
    • ±1.5 cm H2O
  • Volume monitoring:
    • > 210 mL = better than 9%
    • < 210 mL = better than 18 mL
    • < 60 mL = better than 10 mL
  • Pressure monitoring:
    • ±5% or ±2 cm H2O

Alarm Settings

  • Tidal Volume (VTE):
    • Low: OFF, 0 to 1500 mL
    • High: 20 to 1600 mL, OFF
  • Minute Volume (VE):
    • Low: OFF, 0 to 10 L/min
    • High: 0 to 30 L/min, OFF
  • Inspired Oxygen (FlO2):
    • Low: 18 to 100%
    • High: 18 to 100%, OFF
  • Apnea Alarm:
    • Mechanical ventilation ON: > 5 mL breath measured in 30 seconds
    • Mechanical ventilation OFF: > 5 mL breath measured in 30 seconds
  • Low Airway Pressure:
    • 4 cm H2O above PEEP
  • High Pressure:
    • 12 to 100 cm H2O (increments of 1 cm H2O)
  • Sustained Airway Pressure:
    • Mechanical Ventilation ON:
      • PLimit < 30 cm H2O, the sustained limit is 6 cm H2O
      • PLimit 30 to 60 cm H2O, the sustained limit is 20% of PLimit
      • PLimit > 60 cm H2O, the sustained limit is 12 cm H2O
    • PEEP and Mechanical Ventilation ON:
      • Sustained limit increases by PEEP minus 2 cm H2O
    • Mechanical ventilation OFF:
      • PLimit 60 cm H2O, the sustained limit is 50% of PLimit
      • PLimit > 60 cm H2O, the sustained limit is 30 cm H2O
  • Subatmospheric Pressure:
    • Paw < -10 cm H2O
  • Alarm Silence Countdown Timer:
    • 120 to 0 seconds

Ventilator Components

Flow Transducer

  • Type:
    • Variable orifice flow sensor
  • Dimensions:
    • 22 mm OD and 15 mm ID
  • Location:
    • Inspiratory outlet and expiratory inlet
  • Optional autoclavable sensor available

Oxygen Sensor

  • Type:
    • Galvanic fuel cell
  • Life Cycle:
    • Approximately 18 months (dependent on usage)

Anesthetic Agent Delivery

  • Vaporizers:
    • Tec 4, Tec 5, Tec 6 Plus, Tec 7
  • Number of positions:
    • 2 or 3
  • Mounting:
    • Tool-free installation Selectatec® manifold interlocks and isolates vaporizers

Additional information

Brand

GE