Conmed BiCap III Electrosurgical Unit

  • Conmed BiCap III Electrosurgical Unit Electrosurgical Units

BiCap III Electrosurgical Unit (ESU) incorporates ConMed’s Dynamic Response technology, which reacts to different tissue impedances in less than 1 millisecond ensuring consistent delivery of power and clinical effect while minimizing surrounding tissue damage.

Description

Conmed BiCap III Electrosurgical Unit Features

The Conmed BiCap III Electrosurgical Unit (ESU) incorporates ConMed’s Dynamic Response technology, which reacts to different tissue impedances in less than 1 millisecond.

  • ESUm mono and bipolar
  • 200W power
  • Dynamic response technology reacts to different tissue impedances ensuring a consistent delivery of power and clinical effect while minimizing surrounding tissue damage.
  • Front panel volume control
  • Option to select automatic programming which restores the BiCap III to last settings used
  • Pulse Cut Mode provides precise energy bursts for critical dissections

Achieve even energy distribution with the only bipolar hemostasis probe with three pairs of electrodes

  • BiCap Superconductor™ Probes have three pairs of electrodes and a ceramic tip to help provide consistent power distribution and heat dissipation. Other probes have plastic tips with only one pair of electrodes.
  • Omnidirectional electrode spacing and large electrode area provide thermal heating with reduced tissue sticking and larger area coverage.
  • Silver has 35% higher thermal conductivity and 28% less electrical resistivity than gold, resulting in more efficient hemostasis.
  • Available with a short 3.3mm probe tip length for easier passage through a deflected scope.
  • Color-coded shafts allow for the easy identification of size.
  • Central irrigation channel to improve visibility.
  • Conmed BiCap III Electrosurgical Unit Specifications

    Dimensions and Weight

    • Width: 13” (33 cm)
    • Height: 6” (15.44 cm)
    • Depth: 18” (45.72 cm) Including Handle
    • Weight: 14.65 lb. (6.65 kg)

    General

    • Output Control System: Dynamic Response
    • Previous Memory Setting: Yes
    • Contact Quality Monitoring System: (CQM) ARM
    • CQM Range: 10-150 Ω / 30%
    • Cooling System: Convection
    • Self Diagnosis: Yes
    • Technology Isolated: RF Output
    • Duty Cycle: 15 Seconds On, 30 Seconds Off

    Modes

    • Pure Cut
      • Watts: 200
      • Max Peak Voltage: 924V
      • Crest Factor (Typical): 1.4 – 2.0
      • Frequency: 390-410 kHz
    • Blend Cut
      • Watts: 200
      • Max Peak Voltage: 1,200V
      • Crest Factor (Typical): 2.2 – 2.9
      • Frequency: 390-410 kHz

    Special Features – Pulse Cut

    • Coag
      • Watts: 80
      • Max Peak Voltage: 2,560V
      • Crest Factor (Typical): 5.6 – 6.6
      • Frequency: 485-505 kHz
    • Bipolar
      • Watts: 50
      • Max Peak Voltage: 130V
      • Crest Factor (Typical): 1.4 – 2.0
      • Frequency: 390-410 kHz