Bovie IDS 400 Electrosurgical Unit

  • Bovie IDS 400 Electrosurgical Unit

The Bovie IDS 400 has IDS technology that gives the surgeon the flexibility to consistently choose the type of effect desired with maximum convenience and safety. The Bovie IDS Series are innovative electrosurgical generators with fully digital implementation for use in today’s modern OR and surgical outpatient center. The Bovie-IDS units are easy-to-handle yet capable of covering a whole spectrum of electrosurgical interventions. Monopolar and bipolar functions satisfy all surgical demands with maximum safety and a user-friendly interface. Bovie’s digital technology makes controlling the IDS Systems very simple for the OR team, allowing the surgeon to concentrate on what’s important – the operation and the patient.

Description

Bovie IDS 400 Electrosurgical Unit Features

The Bovie IDS 400 Electrosurgical Unit has IDS technology that gives the surgeon the flexibility to consistently choose the type of effect desired with maximum convenience and safety.

  • BovieFDFS (Fast Digital Feedback System) – Digitally controlled CUT and COAG functions give a consistent clinical output.
  • BovieNEM (Neutral Electrode Monitoring)
  • BovieFCFS (First Come First Serve) Improved patient safety with the latest digital implementation.
  • BovieDED (Digital Error Detection)

Bovie IDS 400 Electrosurgical Unit Specifications

Dimensions and Weight

  • Width: 6.5 In (16.5 cm)
  • Height: 14.75 In (37.5 cm)
  • Depth: 18.1 In (46 cm)
  • Weight: < 9.07 kg (< 20 lbs.)

General

  • Duty Cycle: 10 S / 30 S
  • Output Type: RF Isolated “F”
  • Output Frequency: 490 kHZ ± 5
  • Input Current: 4.5 A ~
  • Line Voltage: 100-240 V ~

Operating Parameters

  • Ambient Temperature Range: 10° to 40° C
  • Relative Humidity: 30% to 75%, non-condensing
  • Atmospheric Pressure: 70kPa to 106kPa
  • Warm-up Time: If transported or stored at temperatures outside the operating temperature range, allow one hour for the generator to reach room temperature before use.

Transport and Storage

  • Ambient Temperature Range: -40˚ to +70˚ C
  • Relative Humidity: 10% to 100%, including condensation
  • Atmospheric Pressure: 50kPa to 106kPa

Modes

  • Cut I
    • Output Power: 400 W @ 500Ω
    • Output Frequency: 490 kHz ± 4.9 kHz
    • Crest Factor at Rated Load: 1.6 ± 20%
    • VPeak Max: 1000V
  • Cut II
    • Output Power: 300 W @ 300Ω
    • Output Frequency: 490 kHz ± 4.9 kHz
    • Crest Factor at Rated Load: 1.6 ± 20%
    • VPeak Max: 625V
  • Blend (Max)
    • Output Power: 200 W @ 300Ω
    • Output Frequency: 490 kHz ± 4.9 kHz
    • Repetition Rate: 30 kHz ± 5 kHz
    • Crest Factor at Rated Load: 3.5 ± 20%
    • VPeak Max: 2000V
  • Pinpoint
    • Output Power: 120 W @ 500Ω
    • Output Frequency: 490 kHz ± 4.9 kHz
    • Repetition Rate: 30 kHz ± 5 kHz
    • Crest Factor at Rated Load: 4.5 ± 20%
    • VPeak Max: 2400V
  • Spray
    • Output Power: 80 W @ 500Ω
    • Output Frequency: 490 kHz ± 4.9 kHz
    • Repetition Rate: 30 kHz ± 5 kHz
    • Crest Factor at Rated Load: 6.5 ± 20%
    • VPeak Max: 4000V
  • Bipolar
    • Output Power: 120 W @ 150Ω
    • Output Frequency: 490 kHz ± 4.9 kHz
    • Crest Factor at Rated Load: 1.6 ± 20%
    • VPeak Max: 450V