Baxter 6301 Infusion Pump
The Flo-Gard® 6301 Dual Channel Volumetric Infusion Pump from Baxter can deliver a wide variety of fluids over a broad range of infusion rates. As a Dual Channel pump, it provides the clinician with the convenience of delivering fluids to 2 patients at once. The pump contains 2 linear peristaltic pump heads allowing it to simultaneously infuse 2 different solutions. Each pump head is independently programmable and permits infusion of primary and secondary medication programs. The secondary program automatically switches over to the primary program when a secondary infusion is complete. Upon completion of the primary volume to be infused the pump automatically switches to a keep vein open rate.
The pump uses standard IV administration sets that eliminate the need for costly disposable sets and reduces the potential for waste.
Description
Baxter 6301 Infusion Pump Features
The Baxter 6301 Infusion Pump with Flo-Gard® is a Dual Channel Volumetric Infusion Pump from Baxter and can deliver a wide variety of fluids over a broad range of infusion rates. As a Dual Channel pump, it provides the clinician with the convenience of delivering fluids to 2 patients at once. The pump contains 2 linear peristaltic pump heads allowing it to simultaneously infuse 2 different solutions. Each pump head is independently programmable and permits infusion of primary and secondary medication programs. The secondary program automatically switches over to the primary program when a secondary infusion is complete. Upon completion of the primary volume to be infused the pump automatically switches to a keep vein open rate. The pump uses standard IV administration sets that eliminate the need for costly disposable sets and reduces the potential for waste.
- Dual Volumetric I.V. Pump
- Flow check occlusion alarm
- Flow rate calculation
- Programmed delivery profiles
- Incremental flow rate
- Automatic restart once occlusion clears
- Automatic piggybacking for secondary medications
Baxter 6301 Infusion Pump Specifications
Dimensions and Weight
- Height: 11.4 in (29 cm)
- Width: 13 in (32.5 cm)
- Depth: 5.1 in (13 cm)
- Weight: 11.6 lbs (8.1 kg)
Battery:
- 12 volt, 3.2 sealed lead acid
Battery Life:
- Approximately 6 hours with one pump running at a rate from 1 to 1400 mL/hr using a fully charged battery.
- Approximately 4 hours with both pumps running at a rate from 1 to 1400 mL/hr using a fully charged battery.
- 8 hours for more than 80% recharge.
AC Power Requirements:
- 110/120 V, 60 Hz
Power Cord:
- 2.9 m (9.5 ft) long, with Hospital Grade [plug]
Fuse:
- 0.8 A, 250 V, SB, 5.2 mm (1364 in) x 20 mm (2532 in)
Flow Rate Range:
- Primary Program:
- 1 – 999 mL/hr in 1 mL/hr increments and 1 – 99.9 mL/hr in 0.1 mL/hr increments.
- Upper limit can be reduced by authorized service personnel.
- Secondary Program:
- 1-999 mL/hr in 1 mL/hr increments
- 1 – 99.9 mL/hr in 0.1 ml/hr increments
VTBI FRange
- 1.0- 99.9 ml in 0.1 ml increments of 1-9,999 ml for both primary and secondary of each pump. Upper limit can be reduced by authorized service personnel.
Leakage Current
- Less than 50 micro-amps (per UL 544)
Power Consumption
- 50 W
Air-in-line Detection
- Factory set to NORM, which causes the pump to alarm on air bubbles approximately 75 uL or larger.
- The MIN setting causes the pump to alarm on air bubbles approximately 50 ul or larger.
Baxter 6201 IV Pumps available upon request
Additional information
Brand | Baxter |
---|