Axia V1500A Touch Screen Patient Monitor
The Axia V1500A Touch Screen Patient Monitor is designed for a fast-paced work environment with an intuitive touchscreen. The patient information along with vital sign settings can be quickly modified to meet the needs of a patient’s changing condition. Its upgradeable design fits-in and moves amongst many patient care areas. Offers ECG with arrhythmia detection, standard measurements of non-invasive blood pressure, motion tolerant SpO2, temperature, respiration rate, and capnography.
The Axia V1500A monitor offers several connective solutions to network multiple monitors and/or manage patient data on an electronic medical records platform or a based hospital information system.
Description
Axia V1500A Touch Screen Patient Monitor Features
The Axia V1500A Touch Screen Patient Monitor is designed for a fast-paced work environment with an intuitive touchscreen.
- Optional: integrated Dräger 5-Agent Anesthesia Gas Bench with CO2 capabilities
- High resolution 15 inch touch-screen display
- Capnography with EtCO2 monitoring (optional)
- Auto setting of alarm limits
- Color-coded alarms
- Graphical and tabular trends (72 hours)
- Patient data entry
- OxiCRG, drug dose calculation
- Trace: 8 wave forms
- Trend Time: 1 – 72 Hours
- Networking capability (wireless and non-wireless)
- Color of the waveforms can be changed
- Battery backup (removable batteries)
- Weight: 10.85 lbs.
- Applications: Neonatal, pediatric and adult patients
Axia V1500A Touch Screen Patient Monitor SPECIFICATIONS
GENERAL
- Display
- 12.1 inch color touch screen
- Trace
- 8 waveforms
- Indicator
- Alarm indicator
- Power indicator
- ORS beep and alarm sound
- Trend Time
- 1 – 72 hour
- Recorder: Built-in, thermal array, 3 channels Record width: 48mm
- Recorder paper: 50mm
- Record speed: 25mm/s, 50mm/s
NETWORKING
- Industry standard 802.11 b/g wireless network
POWER
- Source:
- External AC power or internal battery
- AC Power:
- 100 ~ 240VAC, 50/60Hz, 150VA
- Battery:
- Built-in & rechargeable lithium ion
- Operating Time:
- 3+ hours
ENVIRONMENTAL SPECIFICATIONS
- Temperature:
- Operating: 5 ~ 40 °c
- Storage: -20 ~ 65 °c
- Humidity range:
- Operating: s80 %
- Storage: s80 %
ECG
- Input:
- 5-lead ECG cable and standard AAMI line for connection
- Lead Choice:
- I, 11, 111, aVR, aVF, aVL, V, V1-V6, TEST
- Gain Choice:
- x0.5, x1, x2, x4
- Frequency Characteristic:
- 0.05 ~ 35 HZ (+3dB)
- ECG Waveforms:
- 7 channels
- Penetration Voltage:
- 4000VAC 50/60Hz
- Sweep Speed:
- 12.5, 25, 50 and 100 mm/sec (left to right or right to left)
- HR Display Range:
- 30 ~300bpm
- Accuracy:
- ±1bpm or ±1%, whichever is greater
- Alarm Limit Range Setting:
- upper limit 100 ~ 200bpm, lower limit 30 ~ 100bpm
RESP
- Measure Method:
- RA-LL impedance
- Range:
- O ~ 120 rpm
- Accuracy:
- ±3 rpm
- Alarm Limit Setting:
- upper limit 6 ~ 120 rpm
- lower limit 3 ~ 120 rpm
- Sweep Speed:
- 12.5, 25, 50 and 100 mm/sec (left to right or right to left)
SPO2
- ASpO2:
- Anti-motion SpO2
- SpO2% Range:
- 0-100%
- SpO2 Accuracy:
- ±2% (70 ~ 100%, non-motion)
- ±3% (70 ~ 100%, motion)
- Pulse Rate Range:
- 30-250 bpm
- Pulse Rate Accuracy:
- ±2 bpm (non-motion)
- ±3 bpm (motion)
- Alarm Limit Setting:
- upper limit 70 ~ 100%
- lower limit 70 ~ 100%
- SpO2 Probe:
- Red light LED wavelength 660nm±5nm
- Infrared light LED wavelength 940nm±10nm
NIBP
- Measuring Technology:
- Automatic oscillating measurement
- Cuff Inflating:
- <30s (0 ~ 300 mmHg, standard adult cuff)
- Measuring Period:
- AVE <40s
- Mode:
- Manual, Auto
- Measuring Interval in AUTO Mode:
- 2 min~ 4 hrs
- Pulse Rate Range:
- 30 ~ 250 (bpm)
- Adult/Pediatric Mode:
- SYS: 40 ~ 250 (mm Hg)
- DIA :15 ~ 200 (mmHg)
- Neonatal Mode:
- SYS: 40 ~ 135 (mmHg)
- DIA : 15 ~ 100 (mmHg)
- Maximum Mean error:
- ±5mmHg
- Maximum Standard deviation:
- 8mmHg
- Resolution:
- 1mmHg
- Overpressure Protection:
- Adult Mode: 300 (mmHg)
- Neonatal Mode:
- 160 (mmHg)
- Alarm Limit Setting:
- SYS: 50 ~ 240 mmHg
- DIA: 15 ~ 180 mmHg
TEMPERATURE
- Range:
- 25 ~ 50 (°C)
- Accuracy:
- ± 0.2 °c (25.0 ~ 34.9°C)
- ± 0.1 °C (35.0 ~ 39.9°C)
- ± 0.2 °C (40.0 ~ 44.9°C)
- ± 0.3 °C (45.0 ~ 50.0°C)
- Display Resolution:
- 0.1°c
- Alarm Limit Setting:
- Upper limit o ~ 50°C
- Lower limit o ~ 50°C
- Channel:
- 2 channels
IBP
- Measurement Range:
- -50 ~ 300mmHg
- Channel:
- 2 channels
- Pressure Transducer:
- Sensitivity, 5µVN/mmHg
- Impedance Range:
- 300~30000
- Transducer Sites:
- ART, PA,CVP, RAP, LAP, ICP
- Unit:
- mmHg/kPa selectable
- Resolution:
- 1mmHg
- Accurancy:
- ±1 mmHg or ±2%, whichever is greater
- Alarm Range:
- -10 ~ 300mmHg
ETC02
- CO2Measurement Range:
- 0~99mmHg
- Accuracy:
- ±2mmHg (0 ~ 38mmHg)
- 39-99mmHg ±5% of reading +0.08% for every 1 mmHg (above 38mmHg)
- Sampling Rate:
- 50 ml/min
- Initialization Time:
- 30 seconds (typical), reaches ±5% Steady-state accuracy within 3 minutes.
- Respiration Rate:
- 0 ~ 150 breaths
- Mode:
- Adult, Neonate
C.O. (CARDIAC OUTPUT)
- Measurement Method:
- Thermodilution Method
- Measurement Range:
- C.O. 0.1 to 20 Umin
- TB 23 to 43°C
- TI Oto 27 °C
- Resolution:
- C.O. 0.1 Umin
- TB,TI 0.1°C
- Accuracy:
- C.O. ±5% or ±0.1 Umin, whichever is greater, as measured using electronically generated flow curves.
- TB, Tl ±0.1 (without sensor)
- Alarm Range:
- TB 23 to 43°C
- Repeatability:
- C.O. ±2% or ±0.1 Umin, whichever is greater, as measured using electronically generated flow curves.
ANESTHETIC AGENTS
- Method:
- Infrared absorption
- Gas Sorts:
- Halothane, lsoflurane, Enflurane, Sevoflurane, Desflurane, CO2, N2O, 02 (optional Automatic Agent ID)
- Measurement Range:
- Halothane, lsoflurane: 0 ~ 8.5% Enflurane, Sevoflurane: 0 ~ 10% Desflurane: 0 ~ 20%
- CO2: 0 ~ 10%
- N2O: 0 ~ 100%
- 02: 0 ~ 100%
- Bias:
- Halothane, lsoflurane, Enflurane, Sevoflu rane, Desflurane: ±(0.15 Vol%+ 15% rel.)
- CO2: ±(0.5 Vol%+ 12% rel.)
- N2O: ± (2 Vol%+ 8% rel.)
- 02: ±3 Vol%
OTHER STANDARD FEATURES
- OxyCRG
- Drug Dose Calculation
- Cascading ECG
- On screen NIPB trends (up to 250 readings)
- User Set Defaults
- Arrhythmia Detection
- ST Segment
Additional information
Brand | Axia |
---|