Axia V1050T Touch Screen Patient Monitor

  • Axia V1050T Touch Screen Patient Monitor Multiparameter and ECG Monitors

Axia V1050T Touch Screen Patient Monitor is an intuitive touch screen patient monitor; patient information along with vital sign settings can be quickly modified to meet the needs of a patient's changing condition. Its upgradable design fits-in and moves amongst many patient care areas. Offers ECG with arrhythmia detection, standard measurements of non-invasive blood pressure, motion tolerant SpO2, temperature, and respiration rate.

The Axia V1050T monitor offers several connective solutions to network multiple monitors and/or manage patient data on an electronic medical records platform or a based hospital information system.

Product Overview

Description

Axia V1050T Touch Screen Patient Monitor Features

  • High resolution 10.5 inch touch-screen display
  • Auto setting of alarm limits
  • Color-coded alarms
  • Graphical and tabular trends (72 hours)
  • Patient data entry
  • Battery backup (removable batteries)
  • Networking capability (wireless and non-wireless)
  • Color of the waveforms can be changed
  • Applications: Neonatal, pediatric and adult patients

Axia V1050T Touch Screen Patient Monitor OTHER STANDARD FEATURES

  • OxyCRG, drug dose calculation, cascading ECG, On-screen NIPB trends (up to 250 readings), user set defaults, Arrhythmia detection, ST segment

Axia V1050T Touch Screen Patient Monitor Specifications

  • Display: 10.5-inch color touch screen
  • Trace: 8 waveforms
  • Indicator: Alarm indicator, power indicator & QRS beep and alarm sound
  • Networking Capacity: wireless and non-wireless
  • OxyCRG: drug dose calculation
  • Trend Time:
    • 1 – 72 hour
    • Recorder: Built-in, thermal array, 3 channels
    • Record width: 48mm
    • Recorder paper: 50mm
    • Record speed: 25mm/s, 50mm/s
  • Weight: 7.10 lbs.

PERFORMANCE

ECG

  • Input: 5-lead ECG cable and standard AAMI line for connection
  • Lead Choice: I, II, III, aVR, aVF, aVL, V, V1-V6, TEST
  • Gain Choice: x0.5, x1, x2, x4
  • Frequency Characteristic: 0.05 ~ 35 HZ (+3dB)
  • ECG Waveforms: 7 channels
  • Penetration Voltage: 4000VAC 50/60Hz
  • Sweep Speed: 12.5, 25, 50 and 100 mm/sec (left to right or right to left)
  • HR Display Range: 30 ~ 300bpm
  • Accuracy: ±1bpm or ±1%, whichever is greater
  • Alarm Limit Range Setting: upper limit 100

RESP

  • Measure Method: RA-LL impedance
  • Range: 0 ~ 120 rpm
  • Accuracy: ±3 rpm
  • Alarm Limit Setting: upper limit 6 ~ 120 rpm, lower limit 3 ~ 120 rpm
  • Sweep Speed: 12.5, 25, 50 and 100 mm/sec (left to right or right to left)

SPO2

  • ASpO2: Anti-motion SpO2
  • SpO2% Range: 0-100%
  • SpO2 Accuracy: ±2% (70 ~ 100%, non-motion) ±3% (70 ~ 100%, motion)
  • Pulse Rate Range: 30-250 bpm
  • Pulse Rate Accuracy: ±2 bpm (non-motion ±3 bpm (motion)
  • Alarm Limit Setting: upper limit 70 ~ 100%, lower limit 70 ~ 100%
  • SpO2 Probe: Red light LED wavelength 660nm±5nm Infrared light LED wavelength 940nm±10nm

NIBP

  • Measuring Technology: automatic oscillating measurement
  • Cuff Inflating: <30s (0 ~ 300 mmHg, standard adult cuff)
  • Measuring Period: AVE<40s
  • Mode: Manual, Auto
  • Measuring Interval in AUTO Mode: 2 min ~ 4 hrs
  • Pulse Rate Range: 30 ~ 250 (bpm)
  • Adult/Pediatric Mode: SYS: 40 ~ 250 (mmHg) DIA :15 ~ 200 (mmHg)
  • Neonatal Mode: SYS: 40 ~ 135 (mmHg) DIA : 15 ~ 100 (mmHg)
  • Maximum Mean error: ±5mmHg
  • Maximum Standard deviation: 8mmHg
  • Resolution: 1mmHg
  • Overpressure Protection: Adult Mode: 300 (mmHg) Neonatal Mode: 160 (mmHg)
  • Alarm Limit Setting: SYS: 50 ~ 240 mmHg DIA: 15 ~ 180 mmHg

IBP

  • Measurement Range: -50 ~ 300mmHg
  • Channel: 2 channels
  • Pressure Transducer: sensitivity, 5μV/V/mmHg
  • Impedance Range: 300 ~ 3000Ω
  • Transducer Sites: ART, PA,CVP, RAP, LAP, ICP
  • Unit: mmHg/kPa selectable
  • Resolution: 1mmHg
  • Accurancy: ±1mmHg or ±2%, whichever is greater
  • AlarmRange: -10 ~ 300mmHg

TEMP

  • Range: 25 ~ 50 (°C)
  • Accuracy: ± 0.2°C (25.0 ~ 34.9°C) ± 0.1°C (35.0 ~ 39.9°C) ± 0.2°C (40.0 ~ 44.9°C) ± 0.3°C (45.0 ~ 50.0°C)
  • Display Resolution: 0.1°C
  • Alarm Limit Setting: upper limit 0 ~ 50°C, lower limit 0 ~ 50°C
  • Channel: 2 channels

C.O. (CARDIAC OUTPUT)

  • Measurement Method: Thermodilution Method
  • Measurement Range: C.O. 0.1 to 20 L/min TB 23 to 43℃ TI 0 to 27℃
  • Resolution: C.O. 0.1 L/min TB, TI 0.1℃
  • Accuracy: C.O. ±5% or ±0.1 L/min, whichever is greater, as measured using electronically generated flow curves. TB, TI ±0.1(without sensor)
  • Alarm Range: TB 23 to 43℃
  • Repeatability: C.O. ±2% or ±0.1 L/min, whichever is greater, as measured using electronically generated flow curves.

NETWORKING

  • Industry-standard 802.11b/g wireless network

POWER

  • Source: External AC power or internal battery
  • AC Power: 100 ~ 240VAC, 50/60Hz, 150VA
  • Battery: Built-in & rechargeable lithium-ion
  • Operating Time: 3+ hours

ENVIRONMENTAL SPECIFICATIONS

  • Temperature: Operating: 5 ~ 40 °C Storage: -20 ~ 65 °C
  • Humidity range: Operating: ≤80 % Storage: ≤80 %

Additional information

Brand

Axia